ZEISS
8 Troubleshooting | 8.3 Chamber
8.3.2 Defining the Post Initialization Position of the Stage
You can configure the position to which the stage drives after the initialization procedure. Other-
wise, the stage drives to the center position.
Prerequisite
ü
Requires the
Supervisor
privilege.
1. From the Windows start menu, select
SmartSEM > SmartSEM Administrator
.
à
The
SmartSEM Administrator Log on
window is displayed.
2. Enter user name and password.
3. To confirm, click
OK
.
à
The
SmartSEM Administrator
window is displayed showing the user list.
4. Click
Column/Stage
.
5. In the
Stage Post Initialisation Position
input fields, enter the desired position.
Alternatively, use the dual joystick to navigate to the desired position and click
Set to cur-
rent position
.
6. To activate the function, activate the
Post Init. Posn Valid
checkbox.
8.3.3 Changing the Joystick TV Angle
In TV mode (chamberscope), it can occur that dual joystick and stage seem to move to opposite
directions. This is because the selected CCD camera is installed at a certain angle relative to the
stage. Thus, the camera shows a side-inverted view. To remedy this, you need to change the joy-
stick TV angle setting in the software.
Info
If you are working with two CCD cameras: The joystick TV angle can only be set for one CCD
camera. When selecting the other CCD camera, you have to change the setting.
Prerequisite
ü
Requires the
Supervisor
privilege.
1. From the Windows start menu, select
SmartSEM > SmartSEM Administrator
.
à
The
SmartSEM Administrator Log on
window is displayed.
2. Enter user name and password.
3. To confirm, click
OK
.
4. The
SmartSEM Administrator
window is displayed showing the user list.
5. Click
Column/Stage
.
6. In the
Stage Options
section, double-click the
Joystick TV Angle
input field.
7. Enter an angle depending on the installation lo-
cation of the CCD camera.
If the CCD camera is installed at the back, enter
180°
.
If the CCD camera is installed at the front, enter
0°
.
If the CCD camera is installed at the side, enter
90°
.
180°
0°
90°
chamber door
Instruction Manual ZEISS Crossbeam 350 | en-US | Rev. 3 | 349500-8111-000
99