Set-up 

Carl Zeiss 

Connecting the HXP 200 C illuminator 

Fluar Illuminator Z mot. 

 
 

 

14 

435180-9060-701 

05/2012 

 

Switch off the Axio Zoom.V16 microscope and illumination. 

 

Remove the existing tube (Fig. 4/

1

) from the body of the microscope by loosening the hexagon socket 

screw (Fig. 4/

9

) with a ball-head screwdriver (SW 3 mm). 

 

Dismantle the spacer tubes analogously as necessary.  

 

The use of spacer tubes together with the Fluar Illuminator Z mot. is not recommended. 

 

Remove dust caps from Fluar Illuminator Z mot. (Fig. 4/

5

). 

 

Insert the Fluar Illuminator Z mot. into the dovetail mount of the microscope body (Fig.  4/

10

) and 

hand-tighten the hexagon socket screw (Fig. 4/

9

) with the ball-head screwdriver. 

 

The binocular ergo phototube Z (Fig.  4/

1

) should preferably be used for fluorescence 

applications with the Fluar Illuminator Z mot.  
When the ergo phototube Z is assembled for the first time, the shipping screw (hexagon socket 

screw SW 3) on the underside must be screwed out. Keep the shipping screw in a secure place 
for use at a later date. 

 

Insert the tube (Fig. 4/

1

) into the dovetail mount of the Fluar Illuminator Z mot. (Fig. 4/

5

) and hand-

tighten the hexagon socket screw (Fig. 4/

6

) with the ball-head screwdriver. 

 

To prevent eye injury due to scattered light, attach and swing in the anti-glare shield (Fig. 4/

7

on the right or left of the mount or revolving nosepiece (Fig. 4/

8

). 

 
 
 

3.3

 

Connecting the HXP 200 C illuminator  

 

Never pull the light guide out of the Fluar Illuminator when the HXP 200 C is in operation.  

 

Mount the separately delivered HXP 200 lamp in the HXP 200 C illuminator before connecting 
the latter (see manufacturer's instructions). 

 

Remove the protective caps from each end of the HXP light guide (Fig. 4/

3

). 

 

Insert the short end of the HXP light guide (Fig. 4/

3

) until it engages in the light guide socket on the 

reverse side of the Fluar Illuminator Z mot. and secure with clamp screw SW 1 mm (Fig. 4/

2

). 

 

Insert the long end of the HXP light guide (Fig. 4/

3

) into the light guide socket on the reverse side of the 

HXP 200 C illumination unit (Fig. 4/

4

) and push until it engages (locks in automatically). 

 

Use a power cable to connect the power input on the reverse side of the HXP 200 C illuminator with a 

mains outlet. 

 

Summary of Contents for Fluar Illuminator Z mot.

Page 1: ...Operating Manual Fluar Illuminator Z mot for Axio Zoom V16...

Page 2: ......

Page 3: ...d a high quality product from Zeiss Before using the instrument the first time please read this operating manual in order to maintain the high quality of the instrument and ensure reliable work with i...

Page 4: ...ect to updating or revision Unless explicitly authorized dissemination or duplication of this document and commercial exploitation or communication of its contents are not permitted Persons in contrav...

Page 5: ...General information 13 3 2 Assemble the Fluar Illuminator Z mot and connect it up 13 3 3 Connecting the HXP 200 C illuminator 14 3 4 Loading the reflector turret 15 3 5 Inserting or removing the stop...

Page 6: ...al devices they are marked with the mark The operating manual contains all information and warnings the operator must comply with The following warning and information symbols are used in this operati...

Page 7: ...azzling or blinding Never look directly into the light beam or light guide exit The lamp used in the HXP 200 C becomes very hot during operation and has a high internal pressure in a heated state Risk...

Page 8: ...the possible travel path of the motorized table There is a risk of pinching or injury to the fingers when the table is positioned by the control unit Defective instruments are not to be disposed of as...

Page 9: ...arts in particular and improper use The instrument manufacturer shall not be liable for damage caused by faulty operation negligence or any other tampering with the instrument particularly the removal...

Page 10: ...autions and instructions for use must be strictly observed see separate operating manual When operating the microscope always use the anti glare shield to avoid eye damage due to scattered light The c...

Page 11: ...or modules Manual diaphragm slider 423611 0000 000 Motorized diaphragm slider 423612 9901 000 Choice of reflector modules Z FL P C ACR 435180 9160 000 Z FL P C 435180 9260 000 Z FL P C ACR RW 435180 9...

Page 12: ...2 4 Technical data Dimensions Fluar Illuminator Z mot length x width x height 31 mm x 277 mm x 92 mm HXP 200 C illuminator width x height x depth 130 mm x 210 mm x 290 mm Weight Fluar Illuminator Z mo...

Page 13: ...omponents from the packaging and check that all components described on the delivery note are present Remove all shipping braces adhesive tape or similar Keep the original packaging for storage or for...

Page 14: ...for use at a later date Insert the tube Fig 4 1 into the dovetail mount of the Fluar Illuminator Z mot Fig 4 5 and hand tighten the hexagon socket screw Fig 4 6 with the ball head screwdriver To preve...

Page 15: ...number is marked on the right hand side of the module mount Fig 5 3 To install or remove reflector modules Fig 5 5 proceed as follows Installing reflector modules Push the reflector module slightly ti...

Page 16: ...Fluar Illuminator Z mot in the same way However the procedure for removal is different The mounting bay for the diaphragm slider is located on the left hand side of the Fluar Illuminator Z mot Insert...

Page 17: ...e applications producing optical cross sections through the specimen the ApoTome 2 system can be coupled with the Fluar Illuminator Z mot For this purpose the diaphragm shaft is provided with a specia...

Page 18: ...the HXP 200 C illuminator Intensity control dial Fig 9 1 Controls light intensity in stages 0 12 5 25 50 and 100 Shutter button remote open Fig 9 2 Button in pressed position open internal shutter of...

Page 19: ...ch the microscope on by pressing on the HIP After switching on the shutter of the Fluar Illuminator Z mot is always closed Completely open the iris diaphragm of the stop slider man or mot using the co...

Page 20: ...luminator Z mot Fig 8 1 to open the internal shutter integral LED lights up Select the desired magnification on the microscope and adjust the focus as necessary To protect the specimen or reduce scatt...

Page 21: ...or description Error cause Remedy No image visible in eyepieces Microscope not switched on Switch on microscope on HIP Reflector turret cover not closed properly Insert cover correctly and close Switc...

Page 22: ......

Reviews: