Operation
Fluar Illuminator Z mot.
Wearing parts
Carl Zeiss
05/2012
435180-9060-701
21
4.5
Wearing parts
The following wearing parts can be ordered directly from Carl Zeiss:
Designation
Catalogue No.
Remarks
2.5 m CAN-Bus cable
457411-9011-000
HXP light guide with liquid filling, 2m
000000-0482-760
HXP 120 lamp, replacement
000000-0482-730
HXP 200 lamp, replacement
000000-0525-316
4.6
Troubleshooting
Error description
Error cause
Remedy
No image visible in eyepieces
Microscope not switched on
Switch on microscope on HIP
Reflector turret cover not closed
properly
Insert cover correctly and close
Switch microscope on again
Beam path in phototube
switched to camera output
Set beam path to binocular
observation
HXP 200 C illumination unit not
switched on or shutter closed
Switch on HXP 200 C or open
shutter
No image visible in eyepieces
Field stop (stop slider) closed
Open field stop
Reflector turret cannot be turned Reflector modules not accurately
positioned
Remove and re-insert reflector
modules