B-0705-1 01/1992 (Digitized 07/2005)
2-8
blende (2-13/6) visible.
• with the attached at the aperture stop
Set screws centering the aperture.
The mark is a circle on the lens
for orientation.
Then open aperture.
• Two socket wrench 3/65 (2-13/5) from the side
on the centering of the lamp
Fit and the image of the lamp filament
centrally to mark a circle on the
Setting the lens.
• screw securing the lamp housing
(2-13/4) and remove by key B (2-13/3) and
Remove the lamp housing.
• Spiral image using the adjustment screws for
Lamp focusing focusing (2-14/1), if necessary
use the focusing magnifier 6x.
• Not the socket wrench 3/65 (2-14/3) the
Operate adjustment screws for mirrors (2-14/2).
Here, the mirror image of the filament is
moved that image and mirror image about a
Wire thickness laterally offset from each other
are mapped. The result is a homogeneous
luminous surface. Lamp housing
then reattach it.
• The state of adjustment, after the pupil image
Removal of an eyepiece from the
Binocular tube can be controlled.
NOTE:
The lamp voltage is continuously adjustable from
6 to 12 V via the setting dial
(2-13/9) in the control panel LO adjustable.
The current value is on the LED
Indicator "lamp voltage" (2-13/8)
read.
Figure 2-14 Centering the lamp
Figure 2-13 Preparing the
Lamp centering
Page 42
2.2.2.2 Centering of the cross table
• Microscope on.
• Lamp voltage at the control (from 2-15/6) to approximately 9 V
control set value on the LED display (2-15/5).
• Check whether the format plate (see Figure 1-1 / Item 7.4)
into the focusing eyepiece (2-15/8) is used; appropriate, as
Section 2.1.2 (3) use.
Summary of Contents for JENA PLAN HD 9x12
Page 2: ...Original text Page 1 Contribute a better translation Operation instructions JENA PLAN...
Page 9: ...B 0705 1 01 1992 Digitized 07 2005 VI OVERALL VIEW Page 9...
Page 14: ...B 0705 1 01 1992 Digitized 07 2005 1 2 Figure 1 1 microscope equipment JENA PLAN Page 14...
Page 15: ...B 0705 1 01 1992 Digitized 07 2005 1 3 Page 15...
Page 17: ...B 0705 1 01 1992 Digitized 07 2005 1 5 Page 17...
Page 18: ...B 0705 1 01 1992 Digitized 07 2005 1 6 Page 18...
Page 19: ...B 0705 1 01 1992 Digitized 07 2005 1 7 Page 19...
Page 20: ...B 0705 1 01 1992 Digitized 07 2005 1 8 Page 20...
Page 33: ...Page 33 B 0705 1 01 1992 Digitized 07 2005 1 22...
Page 67: ...B 0705 1 01 1992 Digitized 07 2005 2 35 Page 69 3 Care Maintenance Service 3 1 Care...