8  

07/05 

Configuring the beam path and lasers 

• 

Click on the 

LSM

 button in the 

Acquire

 

subordinate toolbar for laser scanning. 

 

Choosing the configuration 

Single Track   

− 

Use for 

single

, double and triple labeling; 

simultaneous scanning only 

− 

Advantage: faster image acquisition 

− 

Disadvantage: cross talk between channels 

Multi Track 

 

− 

Use for double and triple labeling; sequential 
scanning, line by line or frame by frame 

− 

Advantage: when one track is active, only 
one detector and one laser is switched on. 
This reduces cross talk. 

− 

Disadvantage: slower image acquisition 

 

 

• 

Click on the 

Config 

button in the 

Acquire

 subordinate toolbar to 

open the 

Configuration Control

 

window. 

The 

Configuration Control

 window 

appears (Fig. 12). 

 

Setting for single track configuration in Channel Mode 

• 

Select 

Channel Mode

 if necessary (Fig. 12). 

• 

Click on the 

Single Track 

button in the 

Configuration Control

 window (Fig. 12). 

• 

Click on the 

Descanned

 button if necessary. 

The 

Beam Path and Channel Assignment 

panel of the 

Configuration Control 

window displays the 

selected track configuration which is used for the scan procedure.  

• 

You can change the settings of this panel using the following function elements: 

 

Activation / deactivation of the excitation wavelengths (check box) and setting of 
excitation intensities (slider). Open the 

Laser Control 

window via the 

Laser

 button. 

 

Selection of the main dichroic beam splitter (HFT) or secondary dichroic beam splitter 
(NFT) position through selection from the relevant list box. 

 

Selection of an emission filter through selection from the relevant list box. 

 

 

Activation / deactivation (via check box) of the selected channel (Ch 1-4, monitor diode 
ChM, META detectors ChS1-8, transmission ChD) for the scanning procedure by 
assigning an existing color icon or defining a new one.  

 

Fig. 12 

Configuration Control window for 
Single Track 

Summary of Contents for LSM 510 Inverted

Page 1: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 3 5 July 2005...

Page 2: ...e LSM 510 and LSM 510 META laser scanning microscope including its original accessories and compatible accessories from other manufacturers may only be used for the purposes and microscopy techniques...

Page 3: ...nents switch also to ON Now the complete system is ready to be initialized with the LSM Software Switching on the HBO 100 mercury lamp Switch on the HBO 100 mercury lamp via the switch of the power su...

Page 4: ...10 Switchboard window Clicking on this button activates the complete LSM hardware on line mode Click on the Start Expert Mode button in the LSM 510 Switchboard window The LSM 510 Expert Mode Main menu...

Page 5: ...ate toolbar Click on the New button in the File subordinate toolbar The Create New Database window appears Select drive C or D from pull down menu Create a new directory if needed Fig 6 Create New Dat...

Page 6: ...cular tube of the microscope Click on the VIS button to set the microscope for direct observation via the eyepieces of the binocular tube lasers are off Click on the TV button to set the microscope ca...

Page 7: ...ting the microscope for reflected light Click on the Reflected Light button to open the shutter of the HBO 100 mercury lamp Click on the Reflector button and select the desired filter set by clicking...

Page 8: ...ick on the Single Track button in the Configuration Control window Fig 12 Click on the Descanned button if necessary The Beam Path and Channel Assignment panel of the Configuration Control window disp...

Page 9: ...he first line of the Configurations list box and click an Store For loading an existing configuration select it in the list box and click on Apply For deleting an existing configuration select it in t...

Page 10: ...position and diameter When Line button is selected the same rules apply as for Frame Fast Settings Add Track button An additional track is added to the configuration list The maximum of four tracks c...

Page 11: ...lts in the best signal to noise ratio Scan speed 8 usually produces good results Us speed 6 or 7 for superior images Choosing the dynamic range Select the dynamic range 8 or 12 Bit per pixel in the Pi...

Page 12: ...same Optical Size This is important for colocalization studies Image acquisition Once you have set up your parameter as defined in the above section you can acquire a frame image of your specimen Use...

Page 13: ...gh it appears blue on the screen Blue zero minimum Adjusting the laser intensity Set the Pinhole to 1 Airy Unit Fig 21 Set the Detector Gain high When the image is saturated reduce AOTF transmission i...

Page 14: ...upper position of the specimen area where the Z Stack is to start Click on the Mark First button to set the upper position of the Z Stack Then focus on the lower specimen area where the recording of t...

Page 15: ...ing off the system Click on the File button in the Main menu and then click on the Exit button to leave LSM 5 software program Fig 5 If any lasers are still running you should shut them off now in the...

Page 16: ......

Reviews: