LSM 510
LSM 510 - SETUP REQUIREMENTS
LSM 510 META
Scanning Module
Carl Zeiss
B 45-0008 e 10/02
2-15
2.9 Scanning
Module
2 individually driven galvanometric scanners
Scanning speed
Up to ~5 frames/sec (512
×
512 pixels)
Field resolution
Max. 2048
×
2048 pixels (individually adjustable for each axis)
Field of view
10
×
10 mm² with a 1.25
×
objective
Zoom
1
×
... 40
×
, continuous control
Channels
a) Up to 4 channels simultaneously
or
b) 3 traditional confocal channels and 1 META channel
4 confocal reflection/fluorescence channels (PMT)
or
3 PMT and 1 META
1 transmitted light channel (PMT) and 3 NDD
or
4 NDD
1 reference monitor diode
Cooled PMTs (option, forthcoming)
Fiber-optic adaptation of external detectors (option, forthcoming)
Dynamic range
12-bit DAC for each detection channel
Pinholes
4 individual variable pinholes (one per confocal channel)
Computer controlled automatic alignment