QUICKSTART
LSM
510
Carl Zeiss
Quick Start in the Expert Mode
LSM 510 META
4-16
B 45-0008 e 10/02
4.3.3
Turn on the Lasers
This step is used to switch the lasers and set the intensity.
•
Click on the
Laser
button in the
Acquire
subordinate toolbar of the
Main
menu.
−
A
Laser Control
window with a list of
available lasers will appear. Depending on
the lasers available in your system, your
screen may differ from the one displayed.
•
Using the mouse, click on the laser(s) featuring
the appropriate wavelength to excite the dyes
used to label your specimen.
•
In the case of the argon and UV laser
(Enterprise), click on the
Standby
button first.
−
Warming Up
appears in the line
Status
.
−
After 2 minutes, when the warming-up
phase is finished, the
Ready
message
appears.
•
Then click on the
On
button.
−
The laser is switched on.
•
Use
Output [%]
scrollbar to set the intensity to 50 %.
•
In the case of the HeNe laser, click on the
On
button directly.
•
Close the window with the
Close
button.
•
For precise measurements the system should warm up for 2 hours.
•
Depending on the lasers available in your system, your screen may differ from the one displayed.
Fig. 4-14
Laser Control window
Summary of Contents for LSM 510 META
Page 1: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Operating Manual Release 3 2...
Page 8: ...INTRODUCTION LSM 510 Carl Zeiss LSM 510 META VIII B 45 0008 e 10 02...
Page 10: ...NOTES ON DEVICE SAFETY LSM 510 Carl Zeiss Contents LSM 510 META 1 2 B 45 0008 e 10 02...
Page 20: ...NOTES ON DEVICE SAFETY LSM 510 Carl Zeiss LSM 510 META 1 12 B 45 0008 e 10 02...
Page 22: ...LSM 510 SETUP REQUIREMENTS LSM 510 Carl Zeiss Contents LSM 510 META 2 2 B 45 0008 e 10 02...
Page 52: ...QUICKSTART LSM 510 Carl Zeiss Contents LSM 510 META 4 2 B 45 0008 e 10 02...