NOTES ON DEVICE SAFETY
LSM 510
DuoScan
Carl Zeiss
User Interface
LSM 510 META
DuoScan
1-20
B 45-0021 e
03/06
1.11 User
Interface
All user interface ports are equipped with a safety interlock system which warrants laser safety.
These interlock devices must not be manipulated. Other interfaces which are not described here
are service interfaces and are only to be operated by authorized Carl Zeiss service personnel.
The following devices can be mounted and dismounted by the user:
−
Halogen and HBO lamp
−
Transmission PMT
−
Switching Mirror
−
Scan head
1.11.1
Mounting and Dismounting Lamps, TPMT and Switching Mirror
The ports of the lamps, the switching mirror and the transmission PMT are equipped with hardware
interlock devices. At most ports (all ports of the Axioskop 2 FS
MOT
and the Axio Imager.Z1 as well as the
HBO port of the Axiovert 200 M) the following interlock devices are present and have to be operated in
the following way:
Interlock with sensor ring and contact ring:
Fig. 1-9
Sensor ring mounted to the interface
ports on the microscope side
Fig. 1-10
Contact ring mounted to the lamp,
TPMT or switching mirror
The interlock is working when the sensors of the sensor ring (Fig. 1-9/
1
) are depressed by the pins on the
contact ring (Fig. 1-10/
). Whenever this is not the case, for example if the distance between the two
devices is too large, the laser will be blocked and the system cannot be used.