OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Image Optimization
LSM 510 META
DuoScan
4-352
B 45-0021 e
03/06
4.14.1.3
Scan Speed, Scan Average and
Pixel Depth
The signal-to-noise ratio can be substantially
improved by reducing the scanning speed to an
acceptable level and averaging over several scans
(i.e. with an average
Number
greater than 1 for
the
Mean
average
Method
in the
Scan Control
window).
•
Use the
Scan Speed
slider in the
Speed
panel
to set the slowest acceptable scanning speed.
−
The corresponding pixel scanning time (Pixel
Time) and the total scanning time (Scan
Time) are shown in the dialog box.
•
In the
Number
text box of the
Pixel Depth,
Scan Direction & Scan Average
panel enter
the number of measurements to be averaged.
Image optimization can be effected much
faster if you select a smaller frame, since less data
have to be processed.
The greater the number of averages selected for
Mean
average
Method
, the better the image
quality will be; the scanning time will be prolonged
accordingly.
Fig. 4-333
Scan Control window