ANNEX
LSM
510
DuoScan
Carl Zeiss
Monitor Diode
LSM 510 META
DuoScan
8-20
B 45-0021 e
03/06
To use the monitor diode for ratio application,
proceed as follows:
•
Click on the
Ratio
button.
•
Activate the ratio channel R1 or R2 in the
Ratio
Settings
panel in the
Configuration Control
window in addition to the monitor diode
channel (ChM-1) and one PMT channel.
•
Choose the appropriate PMT channel as source
1 in the
Ratio Settings
panel and ChM-1 as
source
2. If this numbering is changed
(inverted), the ratio image will show an
inversion of gray levels if compared to the PMT
image.
It is not possible to do the ratio between
an on-line ratio image generated with
two PMT channels (as in ion-
concentration sensitive ratio imaging) and
the signal of the monitor diode.
Fig. 8-7
Configuration Control window