OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Acquire Menu
LSM 510 META
DuoScan
4-104
B 45-0021 e
03/06
4.5.6
Time Series
The
Time Series Control
window allows the
definition of parameters for time series.
The
Time Series
function offers the following
options for the creation of image series:
•
Definition of break times between 0.1 ms and
10 hours.
•
Determination of the number of steps from 1 to
10,000 for one scanning procedure.
•
Setting of markers.
•
Interruption of time control via pause function,
and resume of the time series function.
•
Triggering of time series via:
−
numeric input
−
external trigger pulses
−
time (of the PC)
4.5.6.1
Open / Close the Time Series
Control Window
•
Click on the
Time Series
button in the
Acquire
subordinate toolbar of the
Main
menu.
The
Time Series Control
window appears on the
screen.
•
Click on the
Close
button to close the
Time
Series Control
window.
The following functions are available on the right-hand side of the
Time Series Control
window:
Close
button
Closes the
Time Series Control
window.
New
button
Opens a new
Image Display
window.
Start T
button
Starts the Time Series.
Start B
button
Starts the Time Series in combination with a bleach procedure. Bleach procedure
must be defined first in the
Bleach Control
window.
Stop
button
Stops the entire Time Series. A current scan is interrupted.
Fig. 4-83
Time Series Control window