LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Acquire Menu
Carl Zeiss
03/06
B 45-0021 e
4-107
4.5.6.3 Stop
Series
Panel
In this panel, the parameters for the end of the
time series are set and the number of cycles is
determined.
The following functions are available:
Manual
button
The time series is finished manually with a click on the
Stop
button.
Trigger
button
The time series is finished via a trigger signal.
Time
button
The time series is finished when the set time has been reached. The internal
computer time applies as the set time.
Number
input box /
arrow keys / slider
Determination of the number of images acquired or image stacks for the
time series.
Time
input box
Input of the time for the end of the time series (
Time
button activated).
Trigger in
list box
Selection of the trigger keys (1-4) with which the end is to be triggered (
Trigger
button activated).
Trigger out
list box Selection of the trigger keys (1-4) for the out signal.
•
Use the slider near
Number
to select the images or image stacks for the time series.
(1)
Stop via Trigger
To end the Time Series via Trigger Control (
Trigger
button activated), first determine the trigger key
which is to end the Time Series.
•
Open the
Trigger in
list box with a click on the
arrow button.
•
Choose one of the trigger keys 1 to 4 (e.g.
Trigger2
).
It is also possible to trigger an out signal via Trigger Control.
•
Open the
Trigger out
list box with a click on the arrow button.
•
Choose one of the trigger keys 1 to 4 (e.g.
Trigger2
).
In this example, the Time Series is ended on pressing key
2
of the Trigger Control, and an out signal is
given at the same time.
Fig. 4-87
Stop Series panel
Fig. 4-88
Stop Series panel