OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Image Optimization
LSM 510 META
DuoScan
4-348
B 45-0021 e
03/06
•
Click on the
Channels
button.
−
This displays the preset parameters of the
configuration loaded.
•
Click on the
Find
button. Make sure to position
the slider correctly. Then scan while the slider is
in the LSM position.
−
This starts the scanning process.
−
The image is seen to build up gradually in a
new window.
Function
Find
produces images of
different brightness for different scan speeds.
As a rule, the first scanned image (Pre-Scan) is not
ideal, since the photomultiplier is not matched to
the light output. More often than not, the screen
image is dull and needs subsequent optimization.
Fig. 4-326
Scan Control window (Channels)
Fig. 4-327
Image Display window