LSM 710 and LSM 780
LEFT TOOL AREA AND HARDWARE CONTROL TOOLS
Systems Maintain
Tab
Carl
Zeiss
02/2010 M60-1-0025
e
145
•
Click the LSM
Continuous
action button to
start a continuous scan. Optimize the pinhole
position in X, Y and Z (Z only for PH1) relative to
the PMT using the
Position X [µm]
,
Position Y
[µm]
and
Position Z [µm]
sliders to maximum
image brightness (see Fig. 157).
•
Moving the
Position [mm]
slider in the
Collimator
panel allows the collimator to be
adjusted to maximum image brightness.
Optimum collimator adjustment obtained in this
way can be stored by clicking on the
Current
Pos
button. The Collimator is pre-set from the
factory and care should be exerted by changing
its position.
•
Click the
Stop
action button to stop the
continuous scan.
Automatic pinhole adjustment:
The automatic adjustment allows the LSM 710 and
LSM 780 systems pinholes to be used with any
combination of beam splitters and more than one
active channel.
Please do not make any program
manipulations while the automatic
pinhole adjustment is running (status
display is red - busy).
•
Click on the
Adjustment
Adjust
Automatically
button.
−
The
Requirements for Adjustment
window will then appear.
•
Meet the requirements listed in the
Requirements for Adjustment
window
(Fig. 158). Press
Cancel
to abort the automatic
pinhole adjustment. Press the
OK
button to
proceed.
−
Pinhole adjustment will then run
automatically and the Scan window will
appear (see Fig.
159). The adjusting
procedure takes approx. 3 min.
−
The determined data are stored
automatically and will be available for all
further examinations using the same
configuration.
Fig. 157
Manual pinhole adjustment
Fig. 158
Requirements for Adjustment window