LSM 710 and LSM 780
CENTER SCREEN AREA / IMAGE CONTAINERS - …
Systems Topography
View
Carl
Zeiss
02/2010 M60-1-0025
e
215
Clicking on the
Color
button opens a color selection box where the color of the
drawing element can be selected with a click of the mouse.
Ticking this box will
hide
all drawing elements, but not the markers.
Sets the
z/x
ratio in the profile diagram to the value
1
. Check: the following creation
of a circle using the relevant tool really results in a circle in the profile display.
Measured angle values correspond to the actual slope of the line displayed.
Tick
in order to see the profile diagram data within a table (below the diagram). A right
mouse button click within the table allows to copy and save this table as a text file (ASCII).
6.12.3.2
Z-histogram Measurement
Chose Z-histogram within the Measure dropdown menu. The Z-histogram will be shown as a diagram
below the image.
6.12.3.3
Bearing Area Ratio Measurement
Chose Bearing area ratio within the
Measure
dropdown menu. The bearing area ratio diagram will be
shown below the image.
Before determination of the top bearing portion, individual peaks (noise, steep slopes) must be
eliminated. The
Median
filter or a suitable longpass filter can be used for this purpose.
Shifting the two cursor crosses within the diagram permits two bearing portions to be given in percent
(e.g. Smr1 = 10 %; Smr2 = 90 %) as default values for which the height difference Rdc is determined
automatically.
6.12.3.4
Roughness Measurement in 2D (Profile) and 3D
•
Chose
Profile
in the
Measure
line to get 2D roughness values.
•
Tick
Roughness
check box:
−
The roughness parameters are calculated and displayed on the right side of the image / profile. All
roughness parameters calculated from a 2D profile are named with
R
, from the 3D surface
RS
.
−
A right mouse click within the roughness value table permits the roughness parameters to be
copied to the clipboard and imported to another program (e.g.: MS Word or MS Excel) via the
Paste