CENTER SCREEN AREA / IMAGE CONTAINERS - …
LSM 710 and LSM 780
Carl Zeiss
Mean of ROI: Additional View Type for Time Series
Systems
222 M60-1-0025
e
02/2010
6.13
Mean of ROI: Additional View Type for Time Series
The
Mean of ROI
View Type allows to
−
display the intensity - time diagram (mean intensity in user defined ROIs over time),
−
use frame - time series and frame - Z-Stack - time series as input,
−
show the intensity values in table form and copy table to clipboard or save as text file.
The
Mean of ROI
function permits interactive
analysis and monitoring intensity over time.
The settings of the
Dimensions, Display
,
Player
and
Overlay
View Options control blocks apply.
The additional view-specific
Mean of ROI View
Option
control block is shown in Fig. 249.
Any changes done with this View Option control
block are effective immediately.
The Image Display in the Mean of ROI View shows
3 panels: the intensity-over-time diagram, the data
table with the intensity values for each ROI over
time and the image display with ROIs in overlay
(see Fig. 250 and Fig. 251).
To access the
Mean of ROI
View Type, load or acquire a time series data set and click on the
Mean ROI
View tab.
The mean intensity of one or several regions of interest is displayed over time. Time flags from bleaching
experiments of trigger events are marked with little symbols on the time-axis of the graph (see Fig. 250).
The
Mean of ROI
analysis always refers to the currently displayed z-section of a Z-Stack (see
Dimension
View Option control block for which section it is).
Fig. 249
Mean of ROI View Option control block