NOTES ON DEVICE SAFETY
LSM 710 and LSM 780
Carl Zeiss
User Interface
Systems
34 M60-1-0025
e
02/2010
2.
The CAN connections on the microscopes are shown in Fig. 28/
2
, Fig. 29/
2
and Fig. 30/
2
. Remove
the connection of the microscope no longer in use at the electronics rack and plug in the connection
of the microscope which should be used instead.
3.
After the connections of the non used microscope are disconnected and the ones of the used
microscope are connected, the system can be switched on again. Before initializing the system with
the LSM Software make sure to use the right database according to the microscope in use. It can be
chosen via the icon
Stand Select
.
1.11.4
Changing the Filter Wheel in the LSM 710 and LSM 780 systems
You can remove and replace filter wheels by push and click.
When opening the cover, the safety control will switch off the laser light.
To exchange the filter wheel of the LSM 710 and LSM 780 systems proceed as follows:
•
Loosen the four fastening screws (Fig. 31/
1
).
•
Remove the cover (Fig. 31/
2
) of the scanning module.
•
Turn the filter wheel (Fig. 32/
1
) until the free position points upwards. Do not touch the filters!
•
Grab the filter wheel on its free position and pull it out upwards.
•
Grab the new filter wheel on its free position. Do not touch the filters!
•
Insert the new filter wheel and snap it into the holder.
•
Close the cover and fasten the four screws.
Fig. 31
Removing the cover
Fig. 32
LSM 710 / LSM 780 filter wheel