CENTER SCREEN AREA / IMAGE CONTAINERS - …
LSM 710 and LSM 780
Carl Zeiss
Polarization Imaging
Systems
266 M60-1-0025
e
02/2010
6.19.1
Light Path for Polarization Imaging
In case emission polarizers (analyzers) are available,
you have a
Notch filter cascade
selection button
available situated on top of the MBS button (see
Fig. 317).
Set up a multitrack configuration using exactly the
same settings with the exceptions of the analyzers.
To this end press the
Notch filter cascade
button.
This opens up the
Polarizer selection box
. Check
the polarizer which you want to use with track 1.
The options are
P
(for the parallel polarized light
transmission) and
S
(for vertical polarized light
transmission) (see Fig. 318).
If you want to ratio image, then it is best for
convenience to choose
P
for track 1. Check the
other polarizer for track 2.
For ratio imaging this would be
S
. For ratio
imaging check the
R1 channel box
and assign
Source 1
to the selected channel-T1 and
Source 2
to the selected channel-T2. You can use the single
PMTs
Ch1
and
Ch2
or the spectral detector
ChS
.
Fig. 318
Light path with Notch filter cascade option
Fig. 317
Light path with Notch filter control
button