MACROS AND VISUAL BASIC
LSM 710 and LSM 780
Carl Zeiss
Macros
Systems
14 M60-1-0025
e
02/2010
Running the Basic Calibration
If you have selected the check box for
Basic Calibration
, please follow these instructions. In
order to run the
Calibration Matrix
only, move on to Running the Calibration Matrix.
•
Make sure that both check boxes have been
selected (Fig. 13)
•
Double check if the objective settings (shown at
the bottom of the macro window) match the
currently used objective
•
Click the
Calibrate
button
•
In order to follow the calibration process, just
click into the
Message
window; a scroll bar will
appear
•
The
Basic Calibration
takes about 15 min
•
The displayed curve should resemble a Gaussian
profile
•
After the macro has finished, a message will
pop up with the request to restart the LSM
software to make the changes effective.
•
Press
Exit
and restart the LSM software
Fig. 13
Collimator Calibration Window