LSM 710 and LSM 780
TOOLS, ADDITIONAL SOFTWARE
Systems
3D for LSM
Carl Zeiss
02/2010 M60-1-0025
e
19
Contrast - Interactive
This function allows interactive changes of the contrast of an image sequence.
Fig. 13
The
Interactive
tab sheet of the
Contrast
dialog window must be selected.
A grey value range of the
Input
image sequence is scaled to another range in the
Output
image
sequence. Both ranges can be edited interactively. This function is used to achieve a better view of an
image sequence, or to scale a range of grey values to single value for a special coding in an image
sequence. The function does not improve the result of the linear segmentation function
Segment
.
Input
indicates the sequence to enhance. If it is a multichannel sequence, a single channel, all channels,
or any number can be selected. The
Input
histogram shows the grey value distribution of the selected
channels of the
Input
image sequence.
Output
defines the name of the result sequence. It will get only those channels which are chosen by the
Input
parameter. The buttons labeled with 8 and 12 define the grey value (intensity) resolution in bit.
Normally the result will get the same resolution as the
Input
sequence. A change will be needed if image
sequences with different resolutions should be combined. Rising the grey value range to 12 bit will not
enhance the display quality or measurement accuracy. The smooth and morphology functions will
produce results with finer gradations.
If
Clip Grey Values
is selected, the output grey values are clipped to the
Low (L)
and
High (H)
values. If
Clip Grey Values
is not selected, output grey values beyond the
Low
and
High
value range are possible.