TOOLS, ADDITIONAL SOFTWARE
LSM 710 and LSM 780
Carl Zeiss
Change Filters
Systems
60 M60-1-0025
e
02/2010
6.2
Change Filters
The
Change Filters
tool is used to update the filter data in the software after a change of filters in the
reflector turret.
•
Close the ZEN software program.
•
Insert the new filter module in the reflector turret.
•
Double-click on the
Change Filters
icon on the desktop.
−
The
Emission Filter & Beam Splitter Control
window appears on the screen. The name of the
currently used database is displayed in the
System
Database
box, with the filter type being
indicated below for checking purposes.
−
The
Filter Cubes Stand
panel shows the
Filter-Wheel No.
and the filter positions available.
−
Use the
Name
and
ID
selection boxes to enter the filters installed in the individual positions of the
filter wheel.
•
Open the
Name
(or
ID
) selection box of the relevant filter position and select the new filter set from
the list.
•
Click on the
Store
button to accept the new settings.
•
Click on the
Close
button to close the
Emission Filter & Beam Splitter Control
window.
All available filter sets have to be registered in the filter list (see
Edit Filter List
function, next
page).
Edit Filter List
The
Edit Filter List
function permits updating of
the filter data in the software after a change of
filters on the stand.
•
Close the ZEN software program.
•
Double-click on the
Change Filters
icon on the
desktop.
•
Click on the
Edit Filter List
button in the
Emission Filter & Beam Splitter Control
window.
−
The
Edit Filter/Beam Splitter List
window
is opened.
This window permits a list of the most frequently used filter sets to be compiled.
•
Click on the arrow button in the
Filtername
list box to open it.
•
Select the filter set which shall be included in the list.
•
Click on the
Apply
button.
The selected filter set is included and displayed in the list (below the
Sumary
list box).
Fig. 41
Edit Filter/Beam Splitter List window