This filter set is now also available in the
Name
selection boxes of the
Filter Cubes Stand
panel and can
be assigned to a filter wheel position.
To remove a filter set which is no longer needed from the list, proceed as follows:
•
Click on the name of the filter set concerned in the list box of the
Edit
Filter/Beam Splitter List
window.
•
Click on the
Remove
button. The filter set is deleted from the list and is then no longer available in
the
Filter Cubes Stand
panel of the
Emission
Filter & Beam Splitter Control
window.
Add New
This function permits new filter sets to be added to
the database.
For this, proceed as follows:
•
Click on the
Add New
button on the
Edit
Filter/Beam Splitter List
window.
−
The
Add New Filter/Beam Splitter
window is opened.
•
Enter the data of the new filter set in the
Filter
Cubes Stand Description
panel, then click on
the
Apply
button.
The new filter set is stored in the database and
included in the
New Filter Cubes Stand
panel.
You can now activate the filter for a filter wheel
position using the procedure described above.
If you have activated the
Non Zeiss
check
box, filter sets from other manufacturers
can also be included in the database.
•
To remove an new filter set from the database, select it with a click of the mouse in the
New Filter
Cubes Stand
panel and then click on
Remove
.
•
Click on
Close
to close the
Add New Filter/Beam Splitter
window.
•
Click on
Close
to close the
Edit Filter/Beam Splitter
List window.
•
Click on the
Store
button to accept the new settings.
•
Click on the
Close
button to close the
Emission Filter & Beam Splitter Control
window.
When you start the ZEN software, the filter data are updated.