TOOLS, ADDITIONAL SOFTWARE
LSM 710 and LSM 780
Carl Zeiss
Maintenance Tool
Systems
64 M60-1-0025
e
02/2010
6.4.2
Starting the Maintenance Tool
Before starting the maintenance tool, the system hardware must be fully activated.
With ZEN 2010, check the following settings:
•
Correct entry of all existing lenses; the calibration objective needs to be entered as "APO Calibration
LSM".
•
For systems with a 405 nm laser, an approximate collimator position should be saved for the
calibration objective.
•
The focus plane needs to be approximately correct
•
All light sources (Hal, LED’s, lasers) need to be activated
•
After start, choose "Log in… Customer".
6.4.3
General operation of the Maintenance Tool
The desired task can be choosen from the list and started with the start button.
The calibration result is displayed with a traffic light type green/yellow/red indicator.
−
If yellow is indicated, repeat the task and try to optimize the settings (e.g. check focus, KÖHLER
lamps, lasers (no standby) etc.
−
If red is indicated, repeat the test one more time; if it still fails, contact your local LSM service
technician.
On LSM 7
DUO
systems, the maintenance tool calibrates only the components of the LSM 710
scanhead, not the LSM 7
LIVE
scanhead.