LSM 710 and LSM 780
TOOLS, ADDITIONAL SOFTWARE
Systems
Maintenance Tool
Carl Zeiss
02/2010 M60-1-0025
e
65
Fig. 47
Maintenance Tool main window
Main beam splitter (MBS) calibration, RGB, InVis:
Allows to automatically adjust the optimal beampath between the MBS, the scanning mirrors and the
master pinhole on all LSM 710 and LSM 780 systems. On LSM 710 and LSM 780 systems with additional
405, 440 or IR (NLO) lasers, make sure the respective lasers are checked.
=> Test tool: calibration objective (turret will be set automatically)
−
KÖHLER illumination for transmitted light must be properly set
The calibration data will be included in the file "… \ bin \ PH1_6.pos".
Scanner calibration
=> Test tool: calibration objective (turret will be set automatically)
−
The scanner calibration requires a properly focused optics (see section 6.4.2).
There is no longer a difference between electrical and optical scanner calibration (as it was with
earlier LSM systems).