LSM 710 and LSM 780
MULTIPHOTON LASER SCANNING MICROSCOPY …
Systems
Non Descanned Detection (NDD)
Carl Zeiss
02/2010 M60-1-0025
e
15
Fig. 15
Configuration tool with NDD tab active. The drop down list shows the available filters. The
one positioned in front of the corresponding channel needs to be selected.
GaAsP nosepiece NDD detector
This detector module is located in a special objective holder. It replaces a standard objective holder on the
Axio Examiner. It holds one objective. Standard objectives need to be used together with the optical
adapter provided.
For attaching the holder to the microscope, check the manual for the Axio Examiner for details on
exchanging objective holders. Make sure the detector is then connected to the real time electronics and
the power supply as described in section 8.2.2.
To indicate the detector to the software use the configuration tool in the following way:
Open the tab for the Microscope and choose the tab Objectives. In the top most drop down list choose
the 1 position Objective nosepiece (Fig. 16).
Fig. 16
Configuration tool with Microscope and Objectives tab active for choosing the objective
nosepiece representing also the GaAsP nosepiece NDD detector.