CONFOCOR
3
Carl Zeiss
System Configuration
ConfoCor 3
28 M60-1-0025
e
02/2010
Measurement / Pinhole Adjustment
•
Press the
Measurement / Pinhole Adjustment
tab to display the laser control panel used for
measurements and pinhole adjustments (Fig. 30).
Press again to close the panel.
•
Activate the lasers you want to use by checking
the corresponding boxes. The lasers with sliders
will appear. Set the requested laser attenuation
(%) using the sliders, the arrows, or typing in
the display field.
Note, that lasers for measurements and pre-
bleach can be selected independently from
each other.
If in
AOTF Dampening
a factor different
from 100 % was chosen, the set AOTF value
in
Measurement / Pinhole Adjustment
will be reduced by that percentage.
Bleach
•
Press the
Bleach
tab to display the laser control
panel used for bleaching (Fig. 31). Press again
to close the panel.
•
Activate the laser or lasers you want to use by
checking the corresponding boxes. The lasers
with sliders will appear. Set the requested laser
attenuation (%) using the sliders, the arrows, or
typing in the display field.
Note, that lasers for measurements and pre-
bleach can be selected independently from
each other.
If in
AOTF Dampening
a factor different
from 100 % was chosen, the set AOTF value
in
Bleach
will be reduced by that
percentage.
The same options are available for the 405 and 440 lasers, if available, in the
Invisible Light
laser
selection box.
All functionalities for laser control are also available in the Laser lines window that will apply for all
selected laser lines (Fig. 32).
Fig. 30
Measurement / Pinhole Adjustment
panel
Fig. 31
Bleach panel