CONFOCOR
3
ConfoCor 3
Save Info
Carl Zeiss
02/2010 M60-1-0025
e
81
9.11
Save Info
The Save Info panel displays all information about
the experiment. It allows you to save all data into a
method.
•
Click on the
Save Info
tab to open the
Save
Info
panel. (Fig. 75). Click again to close the
panel.
•
The
Method Info
window will display all
relevant meta data, including system configu-
ration, measurement protocols and parameter
settings. The scroll bar can be used to view the
complete information.
•
The
Save Infos
window (Fig. 76) will allow you
to save the settings. You can save the method
by pressing the
Save
button. If the method
name exists already, the system will prompt you
to confirm overwriting the old method. Press
Yes
to overwrite,
No
to keep the old method
settings. If the system settings deviating from
default expectations, it will give a warning. For
example, if no pinhole alignment was done or if
a dual color experiment was set up with one
channel or one laser, a corresponding warning
will be displayed. Press
Yes
for continuing
anyway,
No
for returning to the method
without saving.
•
You can enter a new method name or replace
the old one by typing into the
Name
display
window You can enter a category name or
replace an old one by typing into
Category
display box.
•
You can give a comment or replace an old one
by typing it into the
Comment
display box.
•
You can print the content of the
Method Info
window by pressing the
button. This
opens the
Print Setup
window and allows the
parameters of the method to be printed in the
form of a table. You can copy the
Method Info
window into the clipboard by pressing the
Copy
button.
Fig. 75
Save Info panel
Fig. 76
Save Info panel