CONFOCOR
3
ConfoCor 3
Action Buttons
Carl Zeiss
02/2010 M60-1-0025
e
91
The
Table properties
windows (Fig. 89) chosen
from the
Result
context menu, which is opened by
a right mouse click within the table, displays
several options:
−
If the
Show standard deviation
box is
checked, all parameters will be displayed
with their values and the calculated standard
deviation from different measurements.
−
In the
Sort order
display window you can
choose, how different measurements are
grouped together. The hierarchy of grouping
of categories is determined by their listed
sequence.
−
In the
Average
display window you can
select, how the data are averaged. You can
either select
Fit results
, in which case the
single values of the parameters are averaged,
or
Correlation results
, where the single
data points of each correlation functions are
averaged and an average correlation
function is calculated.
Please note, only in the latter case the displayed average correlation function can be meaningfully
fitted.
The
Displayed columns
selection window allows you to change the appearance of the table with regard
to columns to be displayed. All parameters with their boxes checked will be visible in the table. The order
of the displayed columns can be adjusted using drag and drop.
Each row contains a check box on the left-hand side. Deactivation of these check boxes will exclude the
relevant rows from all subsequent evaluation procedures like averaging. These settings will be stored by
saving the data. Immediate reactivation is possible via the check box, if required.
The
Correlation
tool tab contains the following function elements for different displays:
Fig. 89
Table properties window