CONFOCOR
3
Carl Zeiss
Action Buttons
ConfoCor 3
118 M60-1-0025
e
02/2010
the
H
or
L
buttons. The z position can be altered by grapping and sliding the
plane
or by using the arrow
slider. The
plane
shows the actual z-position.
Objective
button: The objective can be selected via graphical pop-up menu. Please note, that for FCS
only the C-Apochromat 40x/60x W N1.2 are specified.
Reflector
button: The reflector cube can be selected via graphical pop-up menu.
Tube Lens
button: The tube lens can be selected via graphical pop-up menu.
HBO Reflected Light
Open/Close Shutter
button: The shutter for reflected epi-fluorescent light from
the HBO lamp is switched on and off by repeated pressing. The light path will be highlighted in color.
For transmitted-light observation
•
Switch the HAL lamp on by the
HAL On/Off
button and adjust the light intensity.
•
Activate the condensor function with the
Condensor
button.
•
Swing the required objective for FCS measurements into the working position. This is performed the
Objective
button.
For Reflected-light Observation (Epi-fluorescence)
•
Turn on the HBO power supply switch.
•
Switch the HBO
Reflected Light Open/Close Shutter
into the Open position.
•
Select the desired filter in the reflector turret using the Reflector button by clicking on it. The filter is
automatically moved into the beam path to enable observation in epi-fluorescence.
•
Select the desired tube lens using the
Tubelens
button by clicking on it. Normally, the correct Lens 1x
will be automatically chosen when the
Ocular
button of the major tool bar is pressed.
•
Swing the required objective for FCS measurements into the working position. This is performed by
selecting the objective with the
Objective
button. We recommend to use the C-Apochromat
40x/1.2 W corr on account of its optimized optics.