80
Operation
G-30-1416-en
OPMI® 1FR XY Surgical Microscope on S7 Floor Stand
Issue 4.0
Printed on 14. 11. 2007
Procedure
•
Switch on the power switch of the suspension system.
•
Set the required brightness of the surgical field illumination on the sus-
pension system.
•
Check the instrument using the checklist.
•
Swing the surgical microscope over the surgical field into an ergonom-
ic position within the working distance.
•
Press the actuator button on the X-Y coupling.
–
The X-Y coupling adopts its center position.
•
Select the lowest magnification using the knob of the magnification
changer.
•
For coarse focusing, look through the eyepieces and lower the surgi-
cal microscope using the suspension arm until the surgical field comes
into focus.
•
Select the highest magnification using the knob of the magnification
changer.
•
Look through the eyepieces and activate the focusing function on the
foot control panel until the microscope is sharply focused on the sur-
gical field.
•
Select the magnification required (magnification changer). Look
through the eyepieces of the binocular tube. Adjust the eyepieces in
such a way that you can see both the edge of the field of view and the
microscope image sharply. Also see “Adjusting the surgical micro-
scope“.
•
Switch off the instrument when you are not using it.