Safety Measures
OPMI LUMERA® 700
Version 7.3
Page 26
G-30-1673-en
9
Switch for manual operation
In the event that one of the major functions fails (XY movement, focus,
zoom, light control), leading to the impairment of further functions, you
can press this switch to change to manual mode.
The system reacts as follows:
– All electrical components except the light source and the lift function
of the ceiling mount are separated from the power supply.
– The light source(s) adjust(s) to a medium brightness level to permit you
to complete the application.
– The 5.7" control panel turns dark.
To be able to continue surgery:
• Manually set the enlargement on the zoom adjustment knob (9.1) of
the microscope.
• Manually position the suspension system for focusing and for moving
the system in the XY direction.
Press switch (9) again to return to normal operation. The switch unlocks
and the system is restarted.
To ensure proper functioning manual mode when required, check its func-
tion at regular intervals (at least every six months) by activating the mode
(without a patient).
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...