System Overview
OPMI LUMERA® 700
Version 7.3
Page 54
G-30-1673-en
3
Slit illuminator (optional)
The slit illuminator is an integratedadditional light source for OPMI
LUMERA 700. It creates a bright, clearly delineated slit image for high-con-
trast viewing of the front and back eye section.
The slit position (right, left) and the size of the slit apertures (slit width
0.2 mm, 2 mm, 3 mm and 4 mm) are set using the 5.7” control panel or
the buttons of the foot control panel or hand grips, depending on the con-
figuration.
The slit height in combination with an f = 200 mm objective lens is
12 mm.
When using the LED light source the slit illuminator is an LED device of
laser class 2 according to IEC 60825-1:2001. According to EN 62471:2008
this light source is to be classified as belonging to the free group.
4
Objective lens
Depending on the scope of delivery the following objective lenses are
available for the different working distances of the surgical microscope:
– Objective lens f = 175 mm (available with and without support ring*)
– Objective lens f = 200 mm (available with and without support ring*)
– Objective lens f = 225 mm (available only with support ring*)
* The support ring is used for mounting the optionally available assistant's
microscope 0°.
When using the LED light source the objective lens is an LED device of laser
class 2 according to IEC 60825-1:2001. According to EN 62471:2008 this
light source is to be classified as belonging to the free group.
LED Class 2
LED Class 2
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...