164
Operation
G-30-1529-en
OPMI® VISU 160 on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 09. 10. 2006
OPMI: setting the lamp brightness
This function permits you to set the lamp brightness.
In the basic mode, the lamp brightness currently set is shown in the upper
display field.
The middle display field remains empty and the associated keys "
∇
" and
"
∆
" have no function.
If the suspension system is equipped with a second halogen lamp
housing, the brightness currently set for the second lamp is displayed in
the middle display field.
Adjusting the settings
Keys "
∇
" and "
∆
" allow you to change the lamp brightness.
–
Check that the lamp brightness can be varied and that brightness var-
iation has an effect on the surgical field illumination. Perform this
check across the entire control range:
–
Halogen: 0.1 ... 1.5
–
Xenon: 0.1 ... 1.8
The brightness can be adjusted in the following ranges:
–
Halogen: 0.1 to 1.5 in steps of 0.1
–
Xenon: 0.1 to 1.8 in steps of 0.1
A beep sounds when brightness level 1.0 has been reached. Level 1.0
corresponds to 500 mW/cm
2
sr, see ISO 10936-2.
Acoustic signals in OPMI VISU
A beep is emitted during brightness adjustment, when level 1.0 has been
reached.
Saving the settings
As soon as you have made a setting, the setting is accepted and saved
under the current user number.
Exception: the brightness setting cannot be saved. It is reset to level 0.1
at the next power-on of the system.
Path: The basic mode is automatically displayed after the instrument has
been switched on.
Basic mode
MODE
STORE MANUAL
USER
1
1
0.7
2
1.5