72
Description
G-30-1529-en
OPMI® VISU 160 on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 09. 10. 2006
Illumination systems
The suspension system is optionally equipped with a Superlux Eye or hal-
ogen illumination system.
1
Superlux Eye illumination
The xenon light source is equipped with an illumination system for
fiber illumination. The xenon lamp generates light whose spectrum re-
sembles that of natural daylight. Regardless of the brightness setting,
the color temperature of the light always remains the same. Normal
daylight film without any additional conversion filters can therefore be
used for photographic documentation. The lamp module contains two
xenon lamps. The second lamp is used as a backup lamp which must
be manually swung into the illumination beam path when the first lamp
fails. You have to pull out the lamp module all the way before being
able to swing in the backup lamp.
Warning!
When using the Superlux Eye illumination system, only operate the
system with special xenon lamps approved by Carl Zeiss for ophthalmic
surgery. If any other than Carl Zeiss-approved xenon lamps are used,
there is the risk of severe injury to the patient's eye.
2
Halogen illumination
The halogen illuminator is equipped with an illumination system for
fiber illumination. The lamp housing contains a backup lamp which is
automatically swung into the illumination beam path when the first
lamp fails.
If required, the illumination system can be equipped with a second
lamp housing so that two separate illumination systems are available
for fiber illumination. The second illumination system can be used, for
example, for a fiber slit lamp or a dual fiber illumination system.
3
Superlux Eye illumination system with additional integrated halogen
illumination (option)
The additional, integrated halogen illumination is a second illumination
system suitable e.g. for the use of a fiber slit lamp or dual fiber illumi-
nation system.