88
Description
G-30-1529-en
OPMI® VISU 160 on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 09. 10. 2006
Identical modules of the suspension systems
Suspension arm
1
Lock of the cable duct
–
For opening, turn a quarter turn clockwise or counterclockwise.
–
For closing, press down and turn a quarter turn clockwise or coun-
terclockwise.
2
Adjustment screw for limiting downward movement
Use this screw to set the minimum vertical working distance from the
surgical field. Bring the surgical microscope into its working position.
Turn the adjustment screw for limiting downward movement clockwise
as far as it will go. Adjust the downward movement limitation before
each surgical procedure.
3
Balance setting screw
After mounting the surgical microscope including all accessories, ad-
just the balance setting of the suspension arm using this screw. Bal-
ance setting is described in detail in chapter “Operation“.
4
Securing screw
for securing the OPMI
®
coupling.
5
Locking knob
for locking the suspension arm in a horizontal position for mounting
the surgical microscope. This prevents the suspension arm from un-
controllably moving upward when insufficient weight is attached.
6
Release bar
Allows non-sterile persons to release the magnetic brakes of the sus-
pension system.
Magnetic brake release buttons
The magnetic brake release buttons are located on the surgical micro-
scope. For as long as you press one of these buttons, you can move the
articulated arm in all directions. When you let go of the button, the mag-
netic brakes will lock all axes in position at the same time.