Failure of magnetic brakes
If the magnetic brakes fail (magnetic brakes are locked), you can manu-
ally position the articulated arm including the microscope by overcoming
the locking effect of the magnetic brakes.
Failure of the X-Y coupling
•
Disconnect the surgical microscope from the suspension system. The
connector is located under the cover of the suspension arm. The illu-
mination remains on.
•
If the X-Y coupling fails, you can manually position the surgical micro-
scope utilizing the movement possibilities of the suspension system
you are using.