4.2.3 Troubleshooting
table
Problem
Cause
Remedy
vignetting or unhomogeneous
image brightness in the field of
view
the field of view is not entirely
visible
The pushrod on the camera tube is
not in the correct position
(observation/photography)
Move the pushrod in the correct
position (end position), see pages
3-11 and 3-13.
Nosepiece with objective not
switched to stop position
Switch nosepiece with objective
in stop position
Condenser not set correctly.
Set condenser correctly
(adjustment, centering), see
pages 3-4, 3-5
Aperture diaphragm not set
correctly.
Set aperture diaphragm correctly
(centering, opening), see page
3-5.
Luminous-field diaphragm not set
correctly.
Set luminous-field diaphragm
correctly (centering, opening), see
page 3-5.
Filter not inserted correctly in filter
mount.
Insert filter correctly in filter
mount, see page 3-3.