background image

8

Safety and Operation 

Device operation safety

The ZEISS UVClean uses UVC radiation to disinfect eyewear.  
UVC radiation is high energy and harmful to living tissue. 
Redundant safety features are implemented into the ZEISS 
UVClean to prevent accidental UVC exposure, however, it is 
important to adhere to the following precautions to ensure 
proper operation:

• The ZEISS UVClean is specifically designed to disinfect 

ophthalmic frames and lenses.  

• 

DO NOT

 pour, spray or use liquids in the device.

• 

DO NOT

 disassemble.

• 

DO NOT

 place device near extreme heat or open 

flames.

• 

DO NOT

 use the ZEISS UVClean to disinfect liquids.

• 

DO NOT

 block air intake on the back of the lid.

Failure to adhere to these instructions may lead to bodily 
injury.

UVC bulb safety

The ZEISS UVClean is pre-calibrated and does not require 
cycle time adjustments.  

UVC bulbs used in the ZEISS UVClean are tested to last 
30,000 cycles, or approximately 3 years, before requiring 
replacement. The air filter should also be replaced at 
the same time as UVC bulb replacement. Instructions on 
replacing the UVC bulbs and air filter are included in the 
ZEISS UVClean Ultraviolet-C (UVC) Eyeglass Disinfection 
Maintenance Kit.

CAUTION: 

UVC exposure on 1.67 hard coat only lenses 

may accelerate the aging process and reduce coating 
lifetime. We always recommend using an AR coating on 
high index lenses.

Ozone safety

While the ZEISS UVClean comes with an internal fan and 
uses bulbs that are designed to not generate ozone, there 
may be a faint smell while the device is running a cycle. 
While sensitive individuals may pick up a slight odor, 
extensive lab testing confirms that ozone levels are well 
below any safety thresholds.

Cleaning device

Interior

To prevent dust buildup within the chamber, always keep 
the ZIESS UVClean lid closed when not in use. 

Never spray 

or pour liquid cleaners inside the device.

Exterior

The device exterior may periodically be wiped with a soft 
cloth and mild cleaner. Never use harsh chemicals or spray 
cleaner directly onto the ZEISS UVClean. Always spray 
cleaner onto soft cloth, then wipe the outside surface of the 
device.

CAUTION: 

Never use liquid cleaner or pour liquids 

directly into the ZEISS UVClean. Severe damage to the 
device or bodily injury may occur.  

Preventative maintenance

The following preventative maintenance tips can help to 
keep your ZEISS UVClean running effectively:

Clean frame holder rails

Frame holder rails should be clean and dust-free to ensure 
optimum UVC illumination and exposure to eyewear. On 
a weekly basis, wipe frame holder rails with a ZEISS Pre-
Moistened Lens Wipe or any lens cleaner sprayed on a soft 
cloth. Never spray lens cleaner directly on rails or inside the 
device chamber.

Remove dust from cleanout tray

A cleanout drawer collects dust in the device and should 
be emptied periodically. Locate the cleanout drawer on the 
back of the ZEISS UVClean. Pull the drawer open and dump 
out any dust or lint accumulated in the tray. A small, dry 
brush may be used to remove collected dust. 

Do not use 

spray or liquid chemicals to clean the drawer.

Recycling and Disposal of used UVC bulbs

UVC bulbs contain trace amounts of mercury and should 
be disposed of according to local regulations for florescent 
lamps. 

Alternatively, used UVC bulbs can be returned for recycling 
if local options are unavailable. Place used UVC bulbs in 
the box from the replacement bulbs and return using the 
mailing label enclosed in the ZEISS UVClean Ultraviolet-C 
(UVC) Eyeglass Disinfection Maintenance Kit.

Cleaning and Recycling

Summary of Contents for UVClean

Page 1: ...ZEISS UVClean User Manual www zeiss com UVClean ...

Page 2: ...Table of Contents 3 General Information 4 Package Contents 5 About the ZEISS UVClean 6 Functionality 8 Safety and Operation 8 Cleaning and Recycling 9 Technical Information ...

Page 3: ... objects as severe damage to the device and or object and injury to the operator may result if misused What it is The ZEISS UVClean Uses UVC light for the safe and effective disinfection of eyeglass frames and lenses Is designed for the unique geometry of eyeglasses to prevent UVC shadows on frames lenses nose pads and temples Direct UVC illumination ensures thorough exposure which is critical for...

Page 4: ...4 Package Contents Device UVC Bulb kit and End caps Quick Start Guide Power cord ...

Page 5: ... the sides and in the middle of the bottom chamber Figure 3 Wearing the provided gloves align the two end pins with the slots on the bulb mounts Figure 4 Push the pins fully into the spring load mounts and twist the bulb turn The pins will click into place Repeat for the remaining three bulbs After installing the bulbs slide each cap outward toward the wall of the chamber Figure 5 Install all four...

Page 6: ...and while LED lights are pulsing blue Eye and skin damage may result if lid is opened while LED lights are pulsing blue LED indicator lights Lights along the underside of the lid illuminate colors Figure 8 and patterns to indicate operating conditions Solid Blue Ready to run a disinfection cycle Pulsing Blue Disinfection cycle running Solid Green Disinfection cycle complete Yellow Bulb replacement...

Page 7: ...ndicator still flashes red proceed to Step 3 Step 3 Replace UVC Bulbs Replace all UVC bulbs according to the instructions included in the ZEISS UVClean Ultraviolet C UVC Eyeglass Disinfection Maintenance Kit To order a ZEISS UVClean Ultraviolet C UVC Eyeglass Disinfection Maintenance Kit please contact your ZEISS Business Development Representative or call Customer Service at 1 800 201 4143 CAUTIO...

Page 8: ...fety thresholds Cleaning device Interior To prevent dust buildup within the chamber always keep the ZIESS UVClean lid closed when not in use Never spray or pour liquid cleaners inside the device Exterior The device exterior may periodically be wiped with a soft cloth and mild cleaner Never use harsh chemicals or spray cleaner directly onto the ZEISS UVClean Always spray cleaner onto soft cloth the...

Page 9: ...isionCare_USA Technical Information Specifications Bulbs 4 non ozone generating UVC lamps Dimensions 12 Wide x 12 High x 20 Deep Weight 20 2 lbs Power cord NEMA 5 15 to IEC 320 C13 for use with 120 volt electrical outlets UVClean electrical power input 120 VAC 50 60 Hz 330 W Listed by NRTL MET Canada USA 250V 10A CAUTION HIGH INTENSITY UV LIGHT DO NOT OPERATE WITH LID OPEN ...