START-UP
VivaTome
Installation Requirements
Carl Zeiss
M60-2-0046 e 05/2012
423639-9000-701
15
3
START-UP
carefully before installing and starting up
the device.
3.1
Installation Requirements
The following requirements apply to the installation of the device:
−
Stable laboratory bench for microscope system on vibration-isolated base plate or vibration-isolating,
air-cushioned microscope stage
Power Requirements:
Two additional wall outlets are required to connect the system (VivaTome and HXP
120).
Ambient Conditions:
1. Operation, specified power
T = 22 °C ±3 °C without interruption (24 h per day,
independent of whether the system is operating or
switched off)
2. Operation, reduced power
T = 10 °C to 35 °C, any conditions different from item 1.
3. Warm-up time
1 h, for high-precision and/or long-term measurements
≥
2 h
4. Relative humidity
< 65 % at 30 °C
5. Operation altitude
max. 2000 m
A constant temperature of 22 °C is ideal for an invariable and optimum performance of the
system.