M U L T I   G A S D E T E C T I O N

6. 

PROGRAMMING

6.2 SETTING DATE FORMAT

This function enables user to choose the correct date format. The default date format is
MM/DD/YYYY (MONTH/DAY/YEAR), as is customary in the USA. To determine to which date
format the instrument is set, enter the supervisory mode (

Section 6.1 Programming Options

)

and use the MODE button to scroll down to “DATE FORMAT”. If set for USA, display will
indicate:

DATE FORMAT

MM/DD

If this format is acceptable, and there is no need to set the date and time, depress the
MODE button to bypass the function. If, however, DD/MM/YYYY (DAY/MONTH/YEAR)
format is desired, depress the ON/OFF button twice, (once to enter the menu, and again to
change the date format) to:

DD/MM

15/06/2000-13:50 (These numbers represent current date & time)

If there is no reason to change the date or time, depress the ON/OFF button a further twelve
times to exit.

6. 

PROGRAMMING

6.3 SETTING DATE/TIME AND DAYLIGHT SAVING

To set date and time enter the supervisory mode (

Section 6.1 Programming Options

) and 

use the MODE button to scroll to “DATE FORMAT”. Depress the ON/OFF button to enter 
the menu. Depress the MODE button. The first digit on the left will start to flash. Depress 
the ON/OFF switch to accept the flashing number and move one number to the right, or 
use the MODE button to change the flashing number to the desired value. If the date format
is MM/DD, the first two digits on the left represent the current month and the first digit must
be either a 0 or a 1. Let us set the date to September 2, 2002. We need 09 to represent the
month of September. If 0 is flashing, depress the ON/OFF switch. This accepts the 0 on 
the left and we may now use the MODE switch to set the next number to 9. Depress the
ON/OFF button to accept the 9 and the month is correctly set to 09. The next digit on the
right is now flashing and the MODE button is used to select 0. Depress the ON/OFF button
to accept the 0, the MODE button to select 2, and the ON/OFF button to accept 2. To set
the year, use the MODE button to select 2, the ON/OFF button to accept the 2. Use the
MODE button to select the 0 and depress the ON/OFF button to accept. Use the MODE
button to select 0 again and depress the ON/OFF button to accept. Use the MODE button to
select the 2 and the ON/OFF button to accept. This completes the date settings. To continue
and set the correct time to 14:45 (2:45PM), use the MODE button to select 1 and the
ON/OFF button to accept the 1. Use the MODE button to select 4 and depress the ON/OFF
button to accept the 4. Use MODE switch to select 4 and depress the ON/OFF switch to
accept the 4. Use MODE to select 5 and ON/OFF to accept and exit this menu.

The date and time are now correctly set. The “DAYLIGHT SAVING” time menu now appears.

DAYLIGHT SAVING

ON

Default is “ON”. Depress the MODE button to accept this setting and exit, or depress the
ON/OFF button to turn on daylight savings and exit.

Summary of Contents for Lumidor MicroMAX+

Page 1: ...M U LT I G A S D E T E C T I O N U S E R A N D M A I N T E N A N C E M A N U A L...

Page 2: ...ty claim is valid Zellweger Analytics will repair or replace the defective product free of charge and send it or any replacement back The decision to repair or replace parts shall be determined by Zel...

Page 3: ...6 4 Auto zeroing 24 TABLE OF CONTENTS SUBJECT SECTION PAGE 6 5 Selecting Pump Mode Diffusion Mode 24 6 6 Selecting Immediate Alarm Mode TWA Alarm Mode 24 6 7 Restoring Factory Defaults 25 6 7 Calibra...

Page 4: ...ems Packaging retain for possible future use Instrument with internal rechargeable battery pack Calibration certificate Standard accessories AC battery charger Alkaline battery holder Dust filter 10 f...

Page 5: ...more than an hour after which time a low battery alarm is indicated by two beeps every 15 seconds Low battery alarm will continue for at least half an hour after which instrument will beep continuous...

Page 6: ...eak reading highest toxic or combustible and lowest oxygen levels STEL 15 minute accumulated short term exposure level and TWA 8 hour time weighted average since instrument was last turned on The STEL...

Page 7: ...bypass each setting depress the MODE button To change any setting depress the ON OFF button For example to enter calibration mode depress the ON OFF button when CALIBRATE is displayed See Section 5 Ca...

Page 8: ...to instrument Place hose in area to be sampled 6 Wait two minutes for full response 7 Take readings 8 Retrieve hose and disconnect from instrument Hose lengths up to 100 feet may be used for pretesti...

Page 9: ...sulfides or silicone containing lubricants present in the atmosphere being monitored The safest possible course of action is to expose the sensors to a known concentration test gas before each day s...

Page 10: ...lay CALIBRATING and count down to 0 7 When CAL COMPLETE is displayed observe which sensors are calibrated and displayed below CAL COMPLETE Remove gas and turn off gas flow 8 Depress the ON OFF switch...

Page 11: ...the menu The following are all built in programmable options Date format date time daylight savings time Auto zero during start up sequence TWA alarm on or off Factory default settings Calibration Pu...

Page 12: ...value If the date format is MM DD the first two digits on the left represent the current month and the first digit must be either a 0 or a 1 Let us set the date to September 2 2002 We need 09 to repre...

Page 13: ...efault values for these alarms are pre set See Appendix A Specifications Section I User Programmable High and Low Alarm Set Points at the factory but are user programmable 2 In the TWA alarm mode thre...

Page 14: ...7 DATA LOGGING 7 4 UNINSTALL PROCEDURE For Windows XP users go to Start Control Panel For other Windows users go to Start Settings Control Panel When the Control Panel opens double click the Add Remo...

Page 15: ...first name and location is 14 14 and 20 alphanumeric characters respectively Downloading at the same date and time for a particular instrument is not allowed For example if you download data from an i...

Page 16: ...LOGGING 7 7 2 GRAPH SCREEN Information on the Graph Screen includes the following See Figure 8 User information employee s name instrument s serial number and user ID Download information date time an...

Page 17: ...on readings Fail indicates a calibration failure and N A indicates that a gas sensor is missing To display the calibration history click on the View menu and then on Calibration History or simply clic...

Page 18: ...ption to OFF The factory alarm setting is immediate alarm with values chosen to provide earlier warning than the STEL and TWA settings The immediate high and immediate low alarms are user programmable...

Page 19: ...TING LOCATION The Edit Location function allows you to modify the location for a specific download After a data file is opened click on the Edit menu then on Edit Location or simply click on the Edit...

Page 20: ...rams such as Microsoft Excel Figure 12 A screenshot of the Import File dialog box Figure 13 A screenshot of the Archive Data dialog box 7 DATA LOGGING 7 13 RECOVERING DATA The Recover Data function en...

Page 21: ...tached to the sensor pins 7 Align pins of new sensor with sockets and push down to secure sensor 8 Reconnect gas plate and secure with screw 9 Reconnect battery connector 10 Install case cover and sec...

Page 22: ...bing Part number GFV194 8 9 SAMPLING HOSE MAINTENANCE Check water trap Part number GFV196 the spherical disc located near the middle of the sampling hose periodically to make sure it is clean and unob...

Page 23: ...nd is the only potentiometer trimpot in the instrument 4 Set digital voltmeter to the DC voltage range black lead to test point 12 analog ground and the red lead to test point 10 These test points are...

Page 24: ...ble 90db at 1 foot Visual Large LED bar Flashing display characters APPENDIX A SPECIFICATIONS G GAS RANGES AND RESOLUTION SENSOR INSTRUMENT RANGE RESOLUTION Combustible LEL 0 100 LEL 1 LEL Oxygen O2 0...

Page 25: ...PPM Hydrogen Sulfide H2S 20PPM 15PPM 10PPM Sulfur Dioxide SO2 10 0PPM 5 0PPM 2 0PPM Phosphine PH3 5 0PPM 1 0PPM 0 3PPM Ammonia NH3 50PPM 35PPM 25PPM Chlorine CL2 5 0PPM 1 0PPM 0 5PPM Chlorine Dioxide...

Page 26: ...10 000 1 Ethylene 500 15 Carbon Monoxide 100 7 Carbon Dioxide 10 000 3 Hydrogen Sulfide 25 25 APPENDIX A SPECIFICATIONS SULPHUR DIOXIDE GAS CONCENTRATION PPM TYPICAL RESPONSE PPM Carbon Monoxide 300pp...

Page 27: ...ermane 1 0 Hydrogen Fluoride 4 0 Alcohols 1000 0 NOTE 1 Short term in minute range APPENDIX A SPECIFICATIONS HYDROGEN CYANIDE GAS CONCENTRATION PPM TYPICAL RESPONSE PPM Hydrogen Sulfide 10 40 Sulfur D...

Page 28: ...Hydrogen Fluoride 3 0 Alcohols 1000 0 Ozone 0 1 0 03 Note Short gas exposure in minute range APPENDIX A SPECIFICATIONS CHLORINE GAS CONCENTRATION PPM TYPICAL RESPONSE PPM Hydrogen Sulfide 10 Neg 0 3 N...

Page 29: ...carbons 10 000 0 Ammonia 30 0 Ozone 1 0 7 Diborane 0 1 0 Arsine 0 1 0 Germane 1 0 Fluorine 3 0 Alcohols 1000 0 Nitrogen Monoxide 100 0 COMBO TOX CO H2S GAS CONCENTRATION PPM TYPICAL RESPONSE PPM Hydro...

Page 30: ...rounding the instrument Refer to Section 8 8 Pump Inlet Filter Replacement for maintenance DO NOT USE non standard sampling tube or fittings Storage Temperature Range 20 degrees C to 50 degrees C 4 de...

Page 31: ...0 P11 52 B Note Max time from Turn On to Ready does not exceed 80 seconds APPENDIX C MODE SEQUENCE STANDARD SEQUENCE DISPLAY LED SPEAKER TO CLEAR PEAK VALUES TO ENTER USER SETUP BEEP 3 BEEPS BEEP BEEP...

Page 32: ...Sequence Standard Mode for how to enter User Set up Mode A DISPLAY LED SPEAKER DEFAULT ON BEEP PUMP STATUS ON APPENDIX D MODE SEQUENCE USER SET UP MODE CALIBRATION LED SPEAKER BEEP 3 BEEPS BEEP OX LEL...

Page 33: ...25 PURGING PLEASE WAIT OX LEL C0 H2S APPLY GAS AUTOMATIC MEMORY CLEAR AFTER PURGING CALIBRATION COMPLETE ONLY SHOWS GASES CALIBRATED NORMAL OPERATING MODE ON OFF YES NO B APPENDIX D MODE SEQUENCE USER...

Page 34: ...P 08 49 B B B FLASH CONTINUOUS HORN FLASH LONG CONTINUOUS BEEPS FLASH LOW GAS ALARM LOW FLOW ALARM 3 BEEPS OX LEL C0 H2S 20 9 0 0 0 DANGER P 08 49 B 3 BEEPS FLASH HIGH GAS ALARM APPENDIX E ALARMS MOD...

Page 35: ...carrying case MCC215A MPRO KIT 1 MINI Kit containing 34 liter calibration gas cylinder GFV243 CH4 CO H2S balance air regulator tubing and water resistant carrying case MCC210A MPRO KIT 2 Kit containi...

Page 36: ...d by the foot GFV256 Tygon hose tubing sold by the foot GFV196 Replacement water filter APPENDIX I REPLACEMENT PARTS PART NO DESCRIPTION 20HFC Replacement dust filter MPLUS DATALOG Data logging softwa...

Page 37: ...first calibration gas is GFV123 the second calibration gas is either GFV 213 58 Liters or GFV 250 34 Liters b MPLUS 4ABCH O2 LEL CO and SO2 Follow the procedure in Section 5 2 Calibration Procedure w...

Page 38: ...M U LT I G A S D E T E C T I O N 9021 Zellweger Analytics 400 Sawgrass Corporate Pkwy Suite 100 Sunrise FL 33325 Tel 1 954 514 2700 Toll free 1 800 538 0363 Fax 1 954 514 2784 MAN MPLUS REV A 0...

Reviews: