25
IFU-001 Rev. 8
15
P
RODUCT
S
PECIFICATIONS AND
T
ECHNICAL
D
ATA
Device Functional Specifications
Parameter
Value
Cold Therapy Temperature Range
Default
34°F, 45°F or 55°F
Custom
34°F - 55°F
Continuous Cycle Length
40°F - 55°F
Heat Therapy Temperature Range
Default
105°F, 107°F or 110 °F
Custom
105°F - 110°F
Continuous Cycle Length
105°F - 107°F
Cycle Length
Default:
10 or 20 minutes
Custom:
3-40 minutes
Continuous Cycle Length:
Active: 10-40 minutes, rest: 30-60 minutes
Edema Pressure Range
Lite
5 mm Hg
Low
20 mm Hg
Medium
45 mm Hg
High
70 mm Hg
DVT Pressure Range
Calf
50 - 70 mm Hg
Foot
90-120 mm Hg
Measured Skin Temperature
Maximum Skin Temperature
Skin temperature measured as high as 107.61
F (42
C) when set to
maximum Heat Reservoir set point for a 40 minute cycle length (110
F)
Minimum Skin Temperature
Skin Temperature measured as low as 40
F (4.44
C) when set to
minimum Cold Reservoir set point for a continuous cycle length (40
F)