MID - Declaration of Conformity
Multidata WR3 is produced and
tested in compliance with the
European Measuring Instruments
Directive (MID). According to this
directive, devices are no longer
carrying an initial verification
stamp, but rather the year of the
device’s declaration of conformity
(recognizable on the front of the
device, for example: M19=2019).
Only valid for Germany:
The metrological part for cooling
energy is calibrated acc. to guide-
line PTB K7.2.
The MID controls the use of mea-
suring devices up to the moment
they are placed on the market
resp. their first putting into use.
After this, the national regulations
for devices subject to compulsory
verification apply within the EU.
The duration of initial verification
validity in Germany remains 5
years for heat meters. After this
period has expired, the measuring
device may no longer be used for
billing in commercial use. The reg-
ulations resp. validity period may
vary in other countries of the EU.
If you have questions, please direct
them to
info@zenner.com
The declaration of conformity is
attached to each measuring instru-
ment. The latest information about
this product can be downloaded
from
www.zenner.com
Safety instructions
Electro-magnetic interference
Multidata WR3 fulfils the national
and international requirements for
interference resistance. To avoid
malfunctions due to other inter-
ferences, do not install fluorescent
lamps, switch cabinets or electric
devices such as motors or pumps
in the immediate vicinity of the
meter. Cables leaving the meter
should not be laid parallel to live
cables (230V), minimum distance
0.2 m.
Care instructions
Clean plastic surfaces with a damp
cloth only. Do not use any scouring
or aggressive cleaning agents! The
device is maintenance-free during
the service life. Repairs may only
be carried out by the manufacturer
or authorized service partners.