How to use the Check Functions on the front panel?
Load the test plan "
DEFAULT_CHECK_TAP.wcf
" with OpenZen-TAP.
The test plan has only 4 activated tests. They all verify one input at the time.
Test 1– AC Clamp Test.
To make the first test:
1) Plug a small wire between the two “Current” connectors.
2) The AC clamp must pass through the wire and be plugged in the input
#1, as seen in the image below.
You are now ready to execute the test.
To go to the test #2, use the right
blue arrow
.
Test 2, 3 and 4 – Accelerometer Tests.
To make the second, third and fourth test, screw an accelerometer to the chassis of the TAP-4 to the
designated area. Then, plug the accelerometer cable into the corresponding input you want to test
and execute the test.