7.16 Adjusting the load display
The load display is for reference only. It is adjusted when new and should not require precise
calibration. The load cell display can be checked as required by adjusting the load adjuster until the
RHGSE just starts to lift off of the ground. If the display is not close to the weight marked on the base
the display can be adjusted by turning the span screw on the display face. Turning the span screw
clockwise increases the displayed reading and counter clockwise decreases the displayed reading.
Once the maximum load (span) is adjusted then the zero displayed reading should be rechecked. This
is accomplished by moving the load lever to the unlocked position and with no load being applied to
the cable, adjust the zero screw to achieve 0000 on the display.
Adjust span screw with load applied
Adjust the zero screw with no load
applied to the load cell