2. DIRECTIONS FOR USE

Step 3:

Turn  bag  upside  down  and 

gently  squeeze. Tilt bag  as 

needed to keep air at tubing 

port (

Figure 2-4

).

Step 4:

Gently pinch orange colored 

tubing 

below

  “ ”  symbol. 

Hold  this  position  until  air  is 

removed from tubing. Gently 

squeeze bag at same time to 

assist fluid flow. If fluid does 

not flow, pinch pressure may 

be too strong. (

Figure 2-5

).

NOTE:  Inside  the  orange 

colored tubing, below the “

” 

symbol is the in-line occlud

-

er. The in-line occluder is the 

built-in  anti  free-flow  valve. 

By  pinching  tubing  gently, 

the tubing moves away from 

the in-line occluder allowing 

fluid to flow (

Figure 2-6

). It is 

important  to  only  pinch  the 

tubing below the “ ” symbol to avoid damaging the in-line occluder.

NOTE: Be extremely careful to pinch tubing using only the flat, soft part of your 

finger. Do not use fingernails when pinching tubing; doing so may puncture the 

delivery set (

Figure 2-7

). If you suspect that the delivery set has been punctured, 

please discard and use a new set

NOTE: Air may also be removed from tubing using the pump prime feature. See 

step 7 for instructions on using the pump prime feature.

Figure 2-4 Squeeze Bag

Figure 2-5 Pinch Tubing

Figure 2-6 Tubing 

Segment Being Pinched

Figure 2-7 Tubing Segment 

Being Punctured by 

Fingernail

Summary of Contents for Infinity Orange

Page 1: ...OPERATOR S MANUAL Small Volume Enteral Feeding Pump...

Page 2: ...osable Set 5 Symbols 6 2 DIRECTIONS FOR USE 7 Priming and Loading the Disposable Set 7 Operating the Pump A Single Feeding Example 10 The Battery 14 Running on Battery Power 14 Charging the Battery 15...

Page 3: ...Infinity Orange Operator s Manual ii...

Page 4: ...inity Orange system includes a complete line of Di 2 ethylhexyl phthalate DEHP free pump delivery sets with automatic free flow protection The Infinity Orange is designed for both hospital and alterna...

Page 5: ...proper air and occlusion sensing and prevent growth of bacteria DO NOT PROGRAM RATE AND DOSE COMBINATIONS WHICH EXCEED A 24 HOUR FEED REGIMEN 4 The battery capacity is an approximation If you are uns...

Page 6: ...ble Sets properly as required by local law 4 If any leaks are detected in the disposable set stop pump operation and disconnect set from patient and replace with a new disposable set 4 Care should be...

Page 7: ...Infinity Orange Operator s Manual vi...

Page 8: ...range Small Volume Enteral Feeding Pump 1 Door 2 Pump Wheel 3 Display 4 Upstream Pressure Sensor 5 Air Sensor 6 Downstream Pressure Sensor 7 Receptacle for Cassette 8 Keypad 1 3 4 5 6 7 8 2 Figure 1 1...

Page 9: ...or s Manual 1 PUMP COMPONENTS Infinity Orange Small Volume Enteral Feeding Pump 1 Pump Instruction Label 2 Speaker 3 Receptacle for Pole Clamp 4 Serial Number Label 5 Charger Power Connector 1 Figure...

Page 10: ...ing Press key again to display total cumulative volume delivered in all feed cycles since total volume was last cleared Press RATE DOSE key to transition between rate and dose settings Press CLEAR key...

Page 11: ...ger is plugged in and connected to a live power outlet The battery symbol indicates pump is running on battery power only The fuel gauge symbol indicates how much battery charge remains Figure 1 8 NOT...

Page 12: ...Order Number INF0500 4EnteraLite Infinity 1200 ml Bag Set Order Number INF1200 4EnteraLite Infinity Spike Set Order Number INF0010 1 Cassette 2 Symbol 3 Downstream Tubing 4 Upstream Tubing 5 100 ml 5...

Page 13: ...on to patient Drop from any angle from height of 3 feet shall not dam age pump operation IEC 529 degree of protection against water entering the enclosure Water jets from any direction shall have no e...

Page 14: ...bag upright and pour in feeding solution Figure 2 2 Close cap securely NOTE Blenderized or aggressively mixed solutions may have foam If using this type of solution allow it to sit for 10 to 15 minute...

Page 15: ...tubing gently the tubing moves away from the in line occluder allowing fluid to flow Figure 2 6 It is important to only pinch the tubing below the symbol to avoid damaging the in line occluder NOTE Be...

Page 16: ...me for one second each The display will then show the letter N followed by a number which is the software revision Next all segments of display will be shown for 2 seconds Verify all display segments...

Page 17: ...bag is at or above the level of the pump door OR If an Infinity brand carry pack is to be used load pump and feeding bag into the proper compartments securing pump bag and tubing with the pack s stra...

Page 18: ...hen the previously programmed dose has been completed and a new feeding is started This counter will also reset to 0 if a feeding is inter rupted and rate or dose is changed or if the pump is turned o...

Page 19: ...pause mode until it is turned off or feeding is restarted To clear DOSE DONE press and hold the ON OFF key for 1 5 seconds to turn pump off or press the RUN PAUSE key to put pump in pause mode Adjustm...

Page 20: ...is turned off and turned back on again the volume delivered coun ter resets to 0 and food type resets to formula Rate dose and food type must be confirmed before restarting the pump with a new feedin...

Page 21: ...pent it will begin to flash indicating approximately 3 hours of charge is left for that bar When the last block is gone the battery symbol will blink to indicate there is approximately 30 minutes of c...

Page 22: ...symbol will appear in the display and the fuel gauge will display four bars scrolling from E to F This pattern continuously repeats while pump is charging Figures 2 18a through 2 18e When bars stop s...

Page 23: ...e 2 21 Backlight To conserve battery life the backlight will turn off automatically 10 seconds after the last key has been pressed However while pump is connected to an A C current the backlight setti...

Page 24: ...O MAX Figure 2 24 To adjust these settings Key Sequence for Adjusting User Preference Settings Step 1 With pump turned off press and hold for 1 5 seconds the and the ON OFF keys at the same time Pump...

Page 25: ...Press the or key to scroll between settings until the desired value is displayed If setting is as desired do not press either key Step 6 Press the PRIME key again Display will read MAX DOSE followed b...

Page 26: ...ER01 ER99 20 3 2 LOAD SET 20 3 3 LOW BATT 20 3 4 NO FLOW IN 21 3 5 NO FLOW OUT 21 3 6 NO FOOD 22 3 7 PUSH RUN TO FEED 22 3 8 SHUT DOOR 23 Messages 3 9 DOSE DONE 23 Indications 3 10 TO STOP LET GO 24 3...

Page 27: ...attempted to run with the set improperly loaded or missing To correct the problem Press the RUN PAUSE key to silence alarm and place pump in pause mode then check the following 1 An Infinity brand dis...

Page 28: ...sors If cleaning is necessary refer to CHAPTER 5 CLEANING Page 29 NO FLOW OUT Figure 3 5 Pump will sound a dual tone beep repeatedly and display will read NO FLOW OUT Why This alarm has occurred becau...

Page 29: ...onnect set from patient press and hold the PRIME key until air bubble has moved to the end of the tubing 3 Check cassette receptacle for cleanliness especially around the air sensor If cleaning is nec...

Page 30: ...This alarm has occurred because the pump door was opened while pump was running To correct the problem Press the RUN PAUSE key to silence alarm and place pump in pause mode then shut the door Check d...

Page 31: ...the key is pressed the display will read MAX When re leased display will read 100 ml hr To correct the problem Change limit in User Pref erence Settings Page 16 CON FIRM DOSE AND FOOD TYPE Figure 3 12...

Page 32: ...the battery is not charging To correct the problem Contact your healthcare provider for service Charger Installed but No Plug Symbol Visible If the plug symbol does not display after the AC Adapter Ch...

Page 33: ...feeding If foam is present in tubing flush by priming pump delivery set either manually or with the pump and resume feeding NO 4 Are air bubbles trapped inside the cassette YES Remove tubing from pum...

Page 34: ...ntact your healthcare provider or ZEVEX Customer Service at 800 970 2337 to receive a new door NO 3 Is pressure sensor region of cassette receptacle clean NO Clean pressure sensor area with a cotton s...

Page 35: ...970 2337 to receive a new charger NO 4 Does the charger port appear to have any damage including bent or missing pins YES Contact your healthcare provider or ZEVEX Customer Service at 800 970 2337 to...

Page 36: ...fectant cleaner Rinse pump by holding under a stream of warm water Then dry with a clean cloth NOTE Avoid harsh cleaners disinfectants They may cause damage to pump surfaces and may affect pump functi...

Page 37: ...gray wheel when pump is in the desired position Tighten black wheel to mount clamp on a pole Figure 6 2 Infinity Brand Adjustable Angle Pole Clamp Order Number 26772 001 Thread bolt into receptacle o...

Page 38: ...er Mini Backpack Order Numbers PCK2001 black PCK2002 light green and gray PCK2003 Specialty Pack call for current colors Pack holds Infinity pump with a 100 ml or 500 ml bag Pump is secured with a Vel...

Page 39: ...a 1200 ml bag Pump is secured with a Velcro strap A Velcro strap secures the neck of the bag The tubing is loaded into pump and any excess length of tubing can be secured with a Velcro tab The downstr...

Page 40: ...LCD OPERATING ORIENTATION Any FLOW RATE Range 0 1 300 ml hr Increment 0 1 ml hr from 0 1 ml hr to 9 9 ml hr 1 ml hr from 10 ml hr to 300 ml hr Accuracy 5 or 050 ml hr whichever is greater see page 40...

Page 41: ...larm NO FOOD Air in tubing bag or set may be empty PUSH RUN TO FEED Pump is on in pause mode for 2 minutes SHUT DOOR Door opened while pump was running ACCESSORIES AC Adapter Charger Multi position Po...

Page 42: ...50 ml hr whichever is greater volumetric flow rate accuracy under the following conditions 8 utilizing ZEVEX Infinity brand disposable sets 8 fluid head height at 6 0 inches 0 3 inches 15 24 cm 0 76 c...

Page 43: ...operating near sources of electromagnetic emissions The Infinity Orange can safely be operated on commercial aircraft and is designed in accordance with EN 60601 1 2 EN 60601 1 4 and RTCA DO160D stan...

Page 44: ...tely 3 5 ml of air is drawn into the tubing before the enteral pump detects bag is empty Therefore approximately 2 5 ml of fluid remains in the disposable set when pump stops feeding Routine Maintenan...

Page 45: ...tion designated by ZEVEX Such repair or replacement shall carry a warranty of ninety 90 days from the date of repair or replacement or the balance of the new product warranty as described above whiche...

Page 46: ...ompanying Documents 6 Figure 1 12 Feeding Bag Connection 6 Figure 1 13 Patient Connection 6 Figure 1 14 TYPE BF Shock Protection 6 Figure 1 15 IEC 529 6 Figure 1 16 Single Patient Use 6 Figure 1 17 DE...

Page 47: ...acklight Settings LITE OFF and LITE ON 16 Figure 2 23 MAX RATE Settings 17 Figure 2 24 MAX DOSE Settings 17 Figure 3 1 ER01 ER99 Alarm 20 Figure 3 2 LOAD SET Alarm 20 Figure 3 3 LOW BATT Alarm 20 Figu...

Page 48: ...w of Internal Straps 31 Figure 6 7 Super Mini Backpack 31 Figure 6 8 Super Mini Backpack Cutaway View of Internal Compartments 31 Figure 6 9 Super Mini Backpack Cutaway View of Internal Straps 31 Figu...

Page 49: ...________ _______________________________________________________________ YOUR PHYSICIAN _______________________________________________________________ ________________________________________________...

Page 50: ...Time 14 Charging the Battery 15 Fuel Gauge 14 LOW BATT Alarm 14 Plug Symbol 15 Running on Battery Power 14 C Carry Packs 31 Cleaning Acceptable Cleaners Disinfectants 29 To clean the AC Adapter Charge...

Page 51: ...4 TO STOP LET GO 9 Instructions for Use Removing Air 7 K Keypad Key 3 Key 3 CLEAR Key 3 FOOD TYPE Key 3 ON OFF Key 3 PRIME Key 3 RATE DOSE Key 3 RUN PAUSE Key 3 VOL TOTAL Key 3 M Maintenance 37 Messag...

Page 52: ...ZEVEX Incorporated 45 U User Preference Settings 16 Adjusting User Preference Settings 17 Alarm Volume 16 Backlight 16 Dose Done 16 Max Dose 17 Max Rate 16 Settings Lock 16 V VOL TOTAL Key 3 11 W Wai...

Page 53: ...Infinity Orange Operator s Manual 46 NOTES...

Page 54: ...6 623 447 6 659 976 6 685 670 6 750 468 6 852 094 6 907 788 6 923 785 D455 489 D501 924 D503 799 D503 978 D504 506 D505 199 and D507 647 and Foreign Patent No s 40108978 9 40209539 1 40302879 5 ZL 01...

Reviews: