Infinity
®
Orange
™
Operator’s Manual
24
3. ALARMS, MESSAGES, AND INDICATIONS
Indications
TO STOP LET GO (Figure 3-10)
When pressing and holding the PRIME key, alarm will
sound once, pump will begin pumping at maximum rate
of 480 ml/hr and display will read “TO STOP LET GO”.
To stop priming:
Release the PRIME key.
MAX (Figure 3-11)
If a rate/dose limit has been programmed in the User
Preference Settings, pressing the
+
key when that limit
has been met will cause the pump to display MAX.
For example, if the User Preference Setting for maxi
-
mum rate has been programmed to 100 ml/hr when
the user attempts to program a rate higher than 100
ml/hr the rate will stop increasing and each time the
+
key is pressed the display will read MAX. When re
-
leased, display will read 100 ml/hr.
To correct the problem:
Change limit in User Pref
-
erence Settings
(Page 16)
.
CON FIRM DOSE AND FOOD TYPE (Figure 3-12)
All feeding parameters must be confirmed before
pump will begin running. This message will display if
the
RUN/PAUSE
key is pressed prior to viewing rate,
dose, and food type parameters.
To correct the problem:
View rate, dose, and
food type. Change settings where required by
using the
+
or
-
key.
NOTE: The food type key allows users to select
whether they are delivering formula (when select
-
ed, screen will read “FORM ULA”) or human milk
(screen will read “HUMN MILK”). The pump auto
-
matically calibrates the air-in-line sensor sensitivity
based on what food type setting is chosen.
Figure 3-10
Priming Indication
Figure 3-11
MAX Indication
Figure 3-12 CON FIRM DOSE