Infinity
®
Orange
™
Operator’s Manual
26
4. TROUBLESHOOTING GUIDE
TROUBLESHOOTING GUIDE
1. Is feeding bag
empty?
YES
Refill bag, re-prime delivery set, and resume feeding.
NO
2. Is air present in the
tubing?
YES
Has all air been removed from
bag
of pump delivery set?
NO
IF YES:
Press and
hold the PRIME key
to move air in the
tubing past pump
delivery set cassette
and resume feeding.
IF NO:
Manually prime tubing by
turning bag upside down, gently
squeezing lower half of bag while
pinching the orange colored tubing
just below the drop symbol. Hold
this position until all air is removed
from bag and tubing.
3. Is formula blender
-
ized or aggressively
mixed?
YES
Let formula sit for 10 - 15 minutes before preparing pump
delivery set for feeding. If foam is present in tubing, flush by
priming pump delivery set either manually or with the pump
and resume feeding.
NO
4. Are air bubbles
trapped inside the
cassette?
YES
Remove tubing from pump and manually prime tubing by
turning bag upside down, gently squeezing lower half of bag
while pinching the orange colored tubing just below the drop
symbol. Hold this position until all air has passed through the
orange segment of tubing.
NO
5. Has the appropriate
food type setting
on the pump been
selected for this
feeding?
NO
Pause pump and change food type by pressing
FOOD TYPE
key and either
+
or
-
to change between formula and human
milk.
YES
If alarm continues, contact your
healthcare provider or ZEVEX Customer Service
at (800) 970-2337 for assistance.
NO FOOD Alarm
If re-priming of the pump delivery set is required, it is advisable to disconnect the
enteral adapter from the patient’s feeding tube while re-priming.