Infinity
®
Orange
™
Operator’s Manual
34
7. SPECIFICATIONS
ALARMS
ER01 - ER99
Self-test error
LOAD SET
Set not loaded properly
LOW BATT
Battery Low
NO FLOW IN
Upstream occlusion
-5 psi (-34 kPa) ± 3psi (21 kPa)
psi (pounds per square inch)
kPa (kilopascals)
NO FLOW OUT
Downstream occlusion,
12 psi (83 kPa) ± 3psi (21 kPa) and 30 sec.
delay before alarm
15 psi (103 kPa) ± 2psi (14 kPa) and 5 sec.
delay before alarm
18 psi (124 kPa) ± 2psi (14 kPa) and 1 sec.
delay before alarm
NO FOOD
Air in tubing - bag or set may be empty
PUSH RUN TO
FEED
Pump is on in ‘pause’ mode for 2 minutes
SHUT DOOR
Door opened while pump was running
ACCESSORIES
AC Adapter/Charger
Multi-position Pole Clamp
Adjustable Angle Pole Clamp
Multi-purpose Infinity brand Custom Carry Packs
WARRANTY
2 years
OPERATING
ENVIRONMENT
Temperature: 41 °F to 104 °F (5 °C to 40 °C)
Humidity: 30% to 95% noncondensing
STORAGE
ENVIRONMENT
Temperature: -4 °F to 149 °F (-20 °C to 65 °C)
Humidity: 10% to 95% noncondensing
CLASSIFICATION
INFORMATION
Infinity Orange Enteral Feeding Pump with internal battery – Type
BF Applied Part
Infinity Orange AC Adapter / Charger – Class II