LIST OF WARNINGS AND CAUTIONS
Warnngs
4
Use ONLY feeding solutions prescribed by the responsible physician, regis
-
tered dietitian, registered nurse, or other licensed practitioner.
4
Use ONLY Infinity brand disposable sets to ensure proper fluid delivery. Oth
-
ers will not deliver the correct dose, may allow dangerous free-flow conditions,
and may generate hazardous fluid pressures which may activate occlusion
alarms at unpredictable pressures.
4
Low flow rates, combined with high dose settings may exceed the life of the
disposable set. The disposable set must be replaced every 24 hours to main
-
tain delivery accuracy, allow proper air and occlusion sensing, and prevent
growth of bacteria. DO NOT PROGRAM RATE AND DOSE COMBINATIONS
WHICH EXCEED A 24 HOUR FEED REGIMEN.
4
The battery capacity is an approximation. If you are unsure that enough ca
-
pacity remains for your intended use, recharge it.
4
To avoid electrical shock, never clean pump with charger plugged into an
outlet or when pump is on.
4
Make sure the Infinity AC Adapter/Charger is completely dry before plugging
into an electrical outlet.
4
Do not use Infinity Orange enteral feeding pump for delivery of non-enteral
solutions. Serious injury may result.
4
If an error occurs (ER01 - ER99 will appear in display), all settings and vol
-
ume counters should be checked and verified before starting a feeding.
4
Prior to starting the pump, confirm rate, dose, and food type are programmed
as prescribed to avoid an unsafe feeding condition.