background image

- 5 - 

test when changing the test ranges. 

7)  To avoid electrical shock or damage to the meter; Do not apply more 

then 500V between any terminal and earth ground. 

8)  To avoid electronic shock, use caution when working above 60VDC or 

25VAC rms, such voltage pose a shock hazard. 

9)  When finished the measurement, switch off the power. Be sure to 

remove the battery when it is not used for a long time to avoid leakage 
problem. 

10)  Do not tamper with the circuitry to avoid damage. 
11)  Do not use or store the instrument in a place of direct sunlight, high 

temperature and high humidity.

 

5. 

METHOD OF MEASUREMENT 

5.1 DCV & ACV MASUREMENT 

1)  Set the Function range switch at the required position. 
2)  Connect black test lead to “COM” terminal and red test lead to the 

“V/OHM” input terminal. 

3)  Connect test leads to measuring point and read the display value the 

polarity of the red lead connection will be indicated at the same time as the 
voltage. 

Note: 

a)  If the voltage to be tested is unknown beforehand, set the 

Function range switch to the highest range and work down. 

b)  When only the figure “1” is displayed over range is being indicated 

and the function range switch has be set to a higher range. 

c)  Never try to measure the voltage above 1000V! Although the 

indication is possible to show, there is danger of damaging the 
internal circuitry. 

5.2 DCA & ACA MEASUREMENT 

1)  Connect the black test lead to the “COM” terminal and the red test lead 

to “A” terminal for a maximum of 0.5A (model DT5801 maximum 2A) 

2)  Set the function range switch at the required position. 
3)  Connect test leads to measuring points and read the display value. The 

polarity at the red test lead connection will be indicated at the same time as 
the current. 

Note: 

a)  If the current range is unknown beforehand, set the function range 

switch to the highest range and work down. 

b)  When only the figure “1” is displayed, over range is being 

Summary of Contents for DT5801

Page 1: ...1 DT58 Series DT5801 DT5802 DT5803 DT5804 DT5805 DT5806 DT5807 DT5808 Large LCD display DIGITAL MULTIMETER OPERATOR S MANUAL...

Page 2: ...ion 5 The is displayed when the battery voltage drops below the operating voltage 6 Full range over load protection 7 Capacitance measurementAuto Zeroing 8 Auto Power Off It will be automatically cut...

Page 3: ...Input impedance 10M Frequency range 40 400Hz 3 DCA DT5803 DT5804 DT5808 DT5801 DT5802 DT5805 DT5806 DT5807 20uA 200uA 2m A 20m A 200m A 2A 20A Range Accuracy 1 8 2 1 8 2 2 0 2 2 0 10 2 0 10 2 0 2 1 8...

Page 4: ...se and connected to the battery snap 2 Don t exceed the input limit shown below Function Range Input terminals Maximum input DCV 200mV 250VDC ACV 200mV 250VAC DCV 2 1000V 1000VDC ACV 2 750V 750VAC OHM...

Page 5: ...e polarity of the red lead connection will be indicated at the same time as the voltage Note a If the voltage to be tested is unknown beforehand set the Function range switch to the highest range and...

Page 6: ...he range selected an over range indication 1 will be displayed and function range switch must be set to a higher range d 200M range has a 10 digits 1M constant the figure will appear in short circuit...

Page 7: ...NUITYTEST 1 Set the function range switch at the position 2 Connect the black test lead to COM terminal and red test lead to V OHM input terminal Note the polarity of the red test lead is 3 This range...

Page 8: ...IS 006P Snap the battery connector leads to the terminals of a new battery and reinsert the battery into the case top Dress the battery leads so that they will not be pinched between the case bottom c...

Reviews: